114 Chapter 8 Maintenance GEInspection Technologies XLG3 VideoProbe System
Replacing the LampTo Determine Whether to Replace the Lamp
If the lamp does not come on when you turn the power switch on, verify all of the following
things before replacing the lamp.
• The probe is properly latched in the handset.
• The lamp was correctly installed.
• The system was not turned on too quickly after being turned off. (Leave the system off
for at least 15 seconds so that it can complete its shutdown sequence. Then turn it
back on.)
• The system is not too hot. (The lamp shuts down in over-temperature conditions.)
• The power switch’s green LED lights up when the switch is on. If not, see “Power
Problems” on page124.
To Remove the Lamp
1. Disconnect all sources of power.
• If using AC power, disconnect the power cord from the power source.
• If a battery is installed, disconnect and remove it. See “To Remove the Battery” on
page 118.
2. Unscrew and remove the lamp access panel.
3. Disconnect the connector above the lamp. Gently remove the lamp.
WARNING LAMP MAY BE HOT. Could cause burns. If the lamp has been
on, wait 10 minutes for it to cool.
Lamp access panel
Connector