3. OPERATING INSTRUCTIONS page 11

Set-Up Waste Streams:

Set-Up Waste streams allows the operator to enter the parameters for the various waste streams.

3.3. NORMAL OPERATION

An operator must be "logged on" before the system can be operated. To log on, the operator must have an operator I.D. and badge number assigned by the system administrator.

To log on, the operator must: press the log on button enter the operator I.D. press <Enter>

enter the badge number press <Enter>

enter the password press <Enter>.

NOTE:

To prevent unauthorized use of the system, the operator must log off by pressing the log off button whenever the unit is left unattended.

After the initial background has been collected and an operator has logged onto the system, the monitor is ready for operation.

Place the container to be scanned on the weight scale platform and close the door to initiate a scan. Note that all detector windows turn gray while the door is open. The monitor will read the scale when the door is closed.

If no weight (<5 lb) is detected in the chamber, the monitor will return to background update mode.

Model BM-286A

Page 14
Image 14
Sigma BM-286A service manual Normal Operation