17. Control by Telnet

Command

Contents

 

 

OFS :

Input DC offset setting

 

Command format

OFS : <1><2>...<33>/[ ]

 

 

<1>

: Measuring unit ID number (0=Recording unit)

 

 

0=Measuring unit 1

1 to 16 CH

 

 

1=Measuring unit 1

17 to 32 CH

 

 

2=Measuring unit 2

33 to 48 CH

 

 

3=Measuring unit 2

49 to 64 CH

 

 

4=Measuring unit 3

65 to 80 CH

 

 

5=Measuring unit 3

81 to 96 CH

 

 

6=Measuring unit 4

97 to 112 CH

 

 

7=Measuring unit 4

113 to 128 CH

 

<2><3>

: CH 1 DC offset

 

 

<4><5>

: CH 2 DC offset

 

 

.

 

 

 

.

 

 

 

.

 

 

<32><33> : CH 16 DC offset

The DC offset value is expressed as a 2-digit hexadecimal number. Ex.) OFS : 000100022334500FFCE2345220034A1FB

00: 0 %

01: 1 %

10: 16 %

32: 50 %

64: 100 % FF : _1 % CE : _50 % 9C : _100 %

Related command : OFS?

OPL :

Output level selection

 

Command format

OPL : <1><2>...<17>

 

 

<1>

: Measuring unit ID number (0=Recording unit)

 

 

0=Measuring unit 1

1 to 16 CH

 

 

1=Measuring unit 1

17 to 32 CH

 

 

2=Measuring unit 2

33 to 48 CH

 

 

3=Measuring unit 2

49 to 64 CH

 

 

4=Measuring unit 3

65 to 80 CH

 

 

5=Measuring unit 3

81 to 96 CH

 

 

6=Measuring unit 4

97 to 112 CH

 

 

7=Measuring unit 4

113 to 128 CH

 

<2>

: CH 1 output level code

 

 

<3>

: CH 2 output level code

 

 

.

 

 

 

.

 

 

 

.

 

 

<17>: CH 16 output level code Output level code

0=RSV, 1=RSV, 2=5 V, 3=2 V, 4=1 V, 5=RSV, 6=RSV, 7=RSV, 8=RSV, 9=RSV, Variable values (Hex) for A=5 or higher.

Related command : OPL?

PNL :

Panel lock

Command format

PNL : <1>/[ ]

 

<1> : 0=Panel lock canceled, 1=Panel lock set

 

Related command : PNL?

RNG :

Input range setting

 

Command format

RNG : <1><2>...<17>/[ ]

 

 

<1>

: Measuring unit ID number

 

 

0=Measuring unit 1

1 to 16 CH

 

 

1=Measuring unit 1

17 to 32 CH

 

 

2=Measuring unit 2

33 to 48 CH

 

 

3=Measuring unit 2

49 to 64 CH

 

 

4=Measuring unit 3

65 to 80 CH

 

 

5=Measuring unit 3

81 to 96 CH

 

 

6=Measuring unit 4

97 to 112 CH

 

 

7=Measuring unit 4

113 to 128 CH

 

<2>

: CH 1 input range code

 

 

<3>

: CH 2 input range code

 

 

.

 

 

 

.

 

 

 

.

 

 

 

<17> : CH 16 input range code

 

 

Input range code

 

 

0=RSV, 1=10 V, 2=5 V, 3=2 V, 4=1 V, 5=0.5 V, 6=0.2 V, 7=0.1 V

 

Related command : RNG?

 

17-6 (E)

SIR-3400H / SIR-3032i / SIR-3032W