17. Control by Telnet

Command

Contents

 

 

RNG?

Input range setting value readout

Command format

RNG? <1><2>/[ ]

 

<1>

: 0=Recording unit setting value, 1=Recording media data

 

<2>

: Measuring unit ID number

 

 

0=Measuring unit 1

 

 

1=Measuring unit 1

 

 

2=Measuring unit 2

 

 

3=Measuring unit 2

 

 

4=Measuring unit 3

 

 

5=Measuring unit 3

 

 

6=Measuring unit 4

 

 

7=Measuring unit 4

Returned format

RNG : <1><2>...<16>

 

<1>

: CH 1 input range code

 

<2>

: CH 2 input range code

 

.

 

 

.

 

 

.

 

 

<16> : CH 16 input range code

 

Input range code

 

 

0=RSV, 1=10 V, 2=5 V, 3=2 V, 4=1 V, 5=0.5 V, 6=0.2 V, 7=0.1 V

 

Related command : RNG

 

 

RPT?

Auto repeat status readout

Command format

RPT? /[ ]

Returned format

RPT : <1>

 

<1> : 0=Other than below

 

 

1=Auto repeat — recording

 

 

2=Auto repeat — rewinding

 

Related command : RPT

 

 

SCH?

Search status

Command format

SCH? /[ ]

Returned format

SCH : <1>

 

<1> : 0=Other than below (after power-on STOP command execution)

 

 

1=Search underway

 

 

2=Search finished (FOUND)

 

 

3=Search finished (NOT FOUND)

 

Related command : SCH

 

 

SLF?

Diagnostics status

Command format

SLF? /[ ]

Returned format

SLF : <1>

 

<1> : 0=Other than below (after power-on STOP command execution)

 

 

1=Diagnostics underway

 

 

2=Diagnostics finished (OK)

 

 

3=Diagnostics finished (NG)

 

Related command : SLF

 

Related command : ERR?

 

 

SPD?

Recording media speed readout

Command format

SPD? <1>/[ ]

 

<1> : 0=Recording unit setting value, 1=Recording media data

Returned format

SPD : <1>

 

<1> : 0=1 x, 1=1/2 x, 2=1/4 x, 3=1/8 x, 4=1/16 x, A=x2, B=x4, F=EXT

 

Related command : SPD

 

 

 

17-14 (E)

SIR-3400H / SIR-3032i / SIR-3032W