SSI America 7SL Calibration Procedure, General Guidelines, Calibration Parameters

Models: 7SL

1 36
Download 36 pages 19.88 Kb
Page 30
Image 30
+0.005% output
+0.001% range
+5 mV
+0.02%
+0.0025 decade better than 0.1°C 0.1 mA AC rms
1 mV
10 m
better than 0.1°C 0.1 mA AC rms

CALIBRATION PROCEDURE

Calibration parameters are logically divided into groups of two parameters each – initial scale value and final scale value. A calibration check is provided after entering the values of each group. It is also possible to perform a calibration check without making an entry: press the FUNC button twice when “OFF” is displayed. The instru- ment goes directly to the group check.

Before beginning calibration, be sure the internal DIP switch V101 is open.

WARNING: Perform the calibration procedure according to J106 jumper positions as shown in Figure 1. Other- wise, the stored calibration values may be lost.

Input Type

 

J106

 

 

1-2

3-4

5-6

7-8

T/C, RTD,

 

 

 

 

0-60 mV, 12-60 mV

close

open

open

open

(standard configuration)

 

 

 

 

0-5 V, 1-5V

open

close

open

open

 

 

 

 

 

0-10V, 2-10 V

open

open

close

open

 

 

 

 

 

0-20 mA, 4-20 mA

open

open

open

close

 

 

 

 

 

Input CTTC R192=180K

 

 

 

SH120=Closed CH101=Open

 

 

 

Figure 1.

J106 Settings

 

 

General Guidelines

a)The instrument should be mounted in its case in order to keep the internal temperature constant.

b)Ambient temperature should be stable. Avoid drift due to air conditioning or other mechanical devices.

c)Relative humidity should not exceed 70%.

d)Minimum warm up time of at least 20 minutes.

e)Operate as much as possible in a noise free environment.

f)during calibration, connect one input at a time to the rear terminal block.

g)Use calibrators with the following:

Accuracy

TC Input:

RTD Input:

CJ Compensation:

Current Transformer:

Resolution

TC Input:

RTD Input:

CJ Compensation:

Current Transformer:

Calibration Parameters

Following is a complete list of calibration symbols: Code Parameter

tL

TC Input Initial Scale Value

tH

TC Input Final Scale Value

t.TC Input Check

rJ

Cold Junction Compensation

rJ.

Cold Junction Compensation Check

PL

RTD Input Initial Scale Value

PH

RTD Input Final Scale Value

P.RTD Input Check

AL

Current Input Initial Scale Value

AH

Current Input Final Scale Value

A.Current Input Check

nL

5

Volt Input Initial Scale Value

nH

5

Volt Input Final Scale Value

n.5 Volt Input Check

UL

10

Volt Input Initial Scale Value

UH

10

Volt Input Final Scale Value

U.10 Volt Input Check

3 0

7sl-6-00.p65

30

9/17/02, 2:44 PM

Page 30
Image 30
SSI America 7SL instruction manual Calibration Procedure, General Guidelines, Calibration Parameters