SSI America 7SL AH Current Input Final Scale Value, Current Input Check, Volt Input Check

Models: 7SL

1 36
Download 36 pages 19.88 Kb
Page 33
Image 33

AL Current Input Initial Scale Value

nL 5 VOLT Input Initial Scale Value

a) Connect calibrator and instrument as shown in

a) Connect calibrator and instrument as shown in

Figure 7.

Figure 8.

10

9

Figure 7. Calibrator Connection

10

9

Figure 8. Calibrator Connection

b)The upper display shows “OFF”, the lower displays shows “AL”.

c)Set calibrator to 0.000 mA (even if the minimum range value is 4mA).

d)Push the button; the display changes to “ON”.

e)After a few seconds, start calibration by pushing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

AH Current Input Final Scale Value

a)Set the calibrator to 20 mA

b)Push the button; the displays will show “ON” and “AH”.

c)After a few seconds, start calibration by pressing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

A.Current Input Check

The display shows “A.” followed by a number showing the measured value in counts.

a)Check the calibration (linear) by setting:

0.000 mA – the readout must be equal to “A.000 00” ± 10 counts;

20.000 mA – the readout must be equal to “A.300 00” ± 10 counts;

10.000 mA – the readout must be equal to “A.150 00” ± 10 counts.

b)Push the FUNC button to go to the next parameter.

b)The upper display shows “OFF”, the lower displays shows “nL”.

c)Set calibrator to 0.000 V (even if the minimum range value is 1V).

d)Push the button; the display changes to “ON”.

e)After a few seconds, start calibration by pushing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

nH 5 Volt Input Final Scale Value

a)Set the calibrator to 5.000 V.

b)Push the button; the displays will show “ON” and “nH”.

c)After a few seconds, start calibration by pressing the FUNC button. When this calibration is complete, the instrument will go to the next parameter.

n.5 Volt Input Check

The display shows “n.” followed by a number showing the measured value in counts.

a)Check the calibration by setting:

0.000 V – the readout must be equal to “n.000 00” ± 10 counts;

5.000 V – the readout must be equal to “n.300 00” ± 10 counts;

2.500 V – the readout must be equal to “n.150 00” ± 10 counts.

b)Push the FUNC button to go to the next parameter.

3 3

7sl-6-00.p65

33

9/17/02, 2:44 PM