2.7Preparation for Storage or Shipping

2.7Preparation for Storage or Shipping

2.7.1Isopropanol Flush

Disconnect the outlet tubing from the pump. Place the inlet filter in isopropanol. Use a syringe to draw a minimum of 50 ml through the pump. Pump a minimum of 5 ml of isopropanol to exit. Leave the inlet tubing connected to the pump. Place the inlet filter in a small plastic bag and attach it to the tubing with a rubber band. Plug the outlet port with the shipping plug or leave a length of outlet tubing on the pump or cover the outlet port with plastic film.

2.7.2 Packaging for Shipping

2.7.1Isopropanol Flush CAUTION: Reship in the original carton, if possible. If the original carton is

2.7.2 Packaging for Shipping not available, wrap the pump in several layers of bubble wrap and cushion the

Manual background bottom, top, and all four sides with 2" of packaging foam. Although heavy, an

Manual background HPLC pump is a delicate instrument and must be carefully packaged to Manual background withstand the shocks and vibration of shipment.

3-6