5.7Lubrication

10.If storing the pump for more than 12 hours fill each pump head with 100% isopropanol then seal the inlets and outlet with the supplied plugs.

5.7Lubrication

The Binary pump has modest lubrication requirements. The bearings in the pump housing and piston carrier are permanently lubricated and require no maintenance. A small dab of light grease such as Lubriplate 630-AA on the cam is the only recommended lubrication. Be sure not to get lubricant on the body of the piston carrier, as this can retard its movement and interfere with proper pumping.

NOTE: Keeping the interior of the pump free of dirt and dust will extend the pump’s useful life.

5.8Other Pump Maintenance

The internal components of the pump can be accessed by removing the pump drawer. To do so, turn off power at front switch. Unplug power cord. Ensure there is no power to the system. Disconnect the inlet line and disconnect pump outlet at the “T” Connector Block above. Unscrew the two thumbscrews on the upper portion of the drawer until fully released (counterclockwise). Pull the drawer out as shown on next page.

Note: Drawer can be removed completely.

7-13