V I . w a r n i n g s f o r c o m p o n e n t s & o p t i o n s

Below are warnings for various standard and optional components for your Xpress. (Options may vary based on your Kid Kart model.) If you fail to heed these warnings, you may cause damage to your Kid Kart, a fall, tip over or a loss of control that may cause severe injury to the child, caregiver or others.

A.TRANSIT OPTION

1.Use only Wheelchair Tiedown and Occupant Restraint Systems (WTORS) which meet the requirements of SAE J2249 Recommended Practice - Wheelchair Tiedown and Occupant Restraint Systems for Use in Motor Vehicles. Do not use WTORS designed to rely on the stroller structure to transfer occupant restraint loads to the vehicle. (See page 11)

2a. The positioning shell/mobility base unit has been dynamically tested with the occupant facing forward mode for a 30 mph (48-km/h) frontal impact test. The occupant must be facing forward during transport if the occupant is greater than 20 pounds in weight.

2b. For infants (20 pounds or less in weight) the positioning shell/mobility base combination must be

facing to the rear. This means the shell is placed in the mobility base so the footbed is positioned away from the handle bar and the occupant is facing the rear of the vehicle.

3.To reduce the potential of injury to vehicle occupants, mobility device mounted accessories, such as trays and respiratory equipment, should be removed and secured separately.

4.Postural supports and positioning devices should not be relied on for occupant restraint.

5.Do not alter or substitute mobility device frame parts, components or seating.

6.Sudden stops or impacts may structurally damage your stroller. Mobility devices involved in such incidents should be replaced.

B.ANTI-TIP TUBES

Anti-tip tubes can help keep your mobility device from tipping. Do not remove the anti-tip tubes and use in down position whenever the base is unfolded.

C.tray supports

1.Do not lift this product by its tray supports.

2.Do not rest drinks on tray supports.

D.CUSHIONS

This mobility device's seating was designed for comfort and postural support. If seating component modifi- cations are necessary, consult your therapist or clinician.