Functional Check
14 P7240 4 GHz 5XAct ive Probe UserManual
NOTE. Now is a good time to perform a probe calibration routine
from your instrument.
6. Disconnectthe probe tip from the oscilloscope, and groundt he
probe tip. (Connect the KlipChip to the probe tip.)
7. With the probeoffset set to 0.0 V,the oscilloscope display should
be at the ground reference.
8. Set the oscilloscopevolts/division to 500 mV.
9. Adjust the probe offset. The displayed waveform should vary
between approximately +1.0 V and --1.0 V.