P7240 4 GHz 5XAct ive Probe UserManual 23
Appendix A: Specifications
These specifications apply to a P7240 probe when used with a
TDS7404 oscilloscope.
The probe and oscilloscopemust first be allowed to warm up for
20 minutes before measurements are taken.

Specifications

CAUTION. Toprevent damage to the probeor circuit under test, do
not apply voltages beyond the nondestructive input voltage range to
the probe.
Table3: Warranted electrical specifications
DC Attenuation Accuracy(probe only) 0.2 ±2% (excludesoffset error)
Output Zero ±10 mV or lessat output of probe
Rise Time(probe only) 120 ps
Table4: Typical electrical characteristics
Bandwidth, (probe only) 4 GHz (See Figure 10)
Linear Input DynamicR ange --2.0V to +2.0 V
(Equivalentto --0.4V to +0.4 V
at the output of the probe.)
Linearity 0.1%over a dynamic range of
--1.75V to + 1.75 V
1.0% over a dynamicrange of
--2.00V to + 2.00 V
NondestructiveInput Voltage Range --30V to +30 V (DC + peak AC)