Teledyne 2230R Serial Communication Commands, Command Summary, Level 0 Commands Description

Models: 2230R

1 26
Download 26 pages 44.62 Kb
Page 16
Image 16
Serial

MODEL 2230R, Process Hydrogen Analyzer

Serial

Communication

Commands

The unit can be communicated with and configured via the use of commands as described below. Two levels of communication outputs are available:

Level 0 – Default level used for data monitoring and basic functions providing a continuous stream of data readings

Level 1 – Password protected level used for configuration of user-settable parameters; interactive single-line data output per command

Command Summary

The RETURN or ENTER key is the last character of the command string. If either key is pressed without a command string the result is an invalid command and will resume continuous display if in Level 0 or return to prompt if in Level 1.

Keystroke

ESC

sp(spacebar)

A

C

=<password>

Level 0 Commands

Description

Stops continuous display to enter a password or command. If in level zero, the continuous display will resume after executing one command.

Pressing the Space key while the serial output is active will display a label line showing the heading for each column of data.

Average readings.

Clear peak hydrogen value.

Enter the password to change security level. A null or invalid password returns to the default security level.

Level 0 password = “0” Level 1 password = “h2scan

Keystroke

A <R1> <R2>

C

D <page>

Level 1 Commands

Description

Set the R1 and R2 set points for the relay contacts in %H.

Clear peak hydrogen value.

Display Product Information. Enter page number 0-6 or A for all pages, default is page 0.

0 – Product information

1 – User configuration

2 – Manufacturing information

3 – Product configuration

4 – Sensor characterization data

5 – Hydrogen calibration data

Page 16

Page 16
Image 16
Teledyne 2230R instruction manual Serial Communication Commands, Command Summary, Keystroke ESC spspacebar A C = password