Teledyne 2230R instruction manual Field Verification Steps, User response, Display

Models: 2230R

1 26
Download 26 pages 44.62 Kb
Page 23
Image 23
Step

MODEL 2230R, Process Hydrogen Analyzer

Step

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

 

FIELD VERIFICATION STEPS

 

 

 

 

Display

 

User response

H2scan:

 

 

Type “ e” to run field verification

 

 

 

 

Ready to Verify

 

 

Type “ y

(Y/N)?

 

 

 

 

 

 

 

Pressure is 1.000

 

 

Type “ n

atm (Y/N)?

 

 

 

 

 

 

 

Enter new pressure:

 

 

Type “ .969” for the local atmospheric pressure

Pressure is 0.9690

 

 

Type “ y

atm (Y/N)?

 

 

 

 

 

 

 

Save as default

 

 

Type “ n

(Y/N)?

 

 

 

 

 

 

 

Verify Sensor

 

 

Type “ y

(Y/N)?

 

 

 

 

 

 

 

Enter Gas (%H2):

 

 

Type “ 1” for gas #1, 1% hydrogen by volume

Settle time (min):

 

 

Type “ 30” for a 30 minute duration for gas #1

 

 

 

 

Tolerance (%H2):

 

 

Type “0.3%” absolute from the for 1% hydrogen by volume

 

 

 

 

Apply 1.000%H2:

 

 

Type “ y

Ready (Y/N)?

 

 

 

 

 

 

Streaming data…

Taking Average…

 

 

An average will be calculated. Verification Gas1 finished

 

 

 

 

Continue test Verify

 

 

Type “ y

Sensor (Y/N)?

 

 

 

 

 

 

 

Enter Gas (%H2):

 

 

Type “ 30” for gas #2, 30% hydrogen by volume

Settle time (min):

 

 

Type “ 30” for a 30 minute duration for gas #2

Tolerance (%H2):

 

 

Type “1.0%” absolute from the TOLERANCE CHART for 30%

 

 

 

hydrogen by volume

 

 

 

 

Apply 30.000%H2:

 

 

Type “ y

Ready (Y/N)?

 

 

 

 

 

 

 

Streaming data…

 

 

 

 

 

 

 

18

Taking Average…

An average will be calculated. Verification Gas2 finished.

 

19Continue test Verify Type “n” to end field verification.

 

 

Sensor (Y/N)?

Verification complete.

4).

If the unit passes Verification, Calibration is not required at this time.

5).

If the unit fails Verification, you will see the following:

 

 

 

 

18

 

Taking Average…

An average will be calculated.

19

 

Verify Failed

Verification complete.

6).

If the unit fails Verification, TELEDYNE ANALYTICAL INSTRUMENTS recommends

 

Calibration, detailed in the following section.

Page 23

Page 23
Image 23
Teledyne instruction manual Field Verification Steps, User response, MODEL 2230R, Process Hydrogen Analyzer, Display