Teledyne instruction manual MODEL 2230R, Process Hydrogen Analyzer, Display, User Response

Models: 2230R

1 26
Download 26 pages 44.62 Kb
Page 19
Image 19
two place hexadecimal value:

MODEL 2230R, Process Hydrogen Analyzer

Now use the 4 BIT-TO-HEXADECIMALtable above to convert this 4 bit value combination into a

two place hexadecimal value:

 

<fmt> Two Place Hexadecimal Value :

8b

From the OPTIONS <opt> table above, you identify your desired columns with its corresponding bit value:

<opt> Descriptions

<opt> Bit Value

Calculations Errors

0000 0100

Heater State

0000 0010

<opt> 4 Bit Value Combination:

0000 0110

Again, use the 4 BIT-TO-HEXADECIMALtable to convert this 4 bit value combination into a two place hexadecimal value:

<opt> Two Place Hexadecimal Value :

06

Conclusion: To have the Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak Hydrogen Reading, Calculations Errors and Heater State columns continuously reported, you will implement the “G” serial command (“Go” Command, refer to LEVEL 1 COMMANDS table) as follows at the H2scan:

command prompt: G < fmt> <opt> = g 8b 06

Display

User Response

H2scan:

Type “g 8b 06”

Page 19

Page 19
Image 19
Teledyne instruction manual MODEL 2230R, Process Hydrogen Analyzer, Display, User Response, two place hexadecimal value