MODEL 2230R, Process Hydrogen Analyzer

3. OPERATION

3.1UNIT LOCATION

The unit can be mounted in any orientation or position, however vertical mounting (unit above process connection) should be made in process streams containing liquids or potentially condensing gases.

3.1.1 WARNING LABEL

Remove sensor tube warning label prior to using the unit.

3.2MOUNTING

Mounting is achieved by securing the sensor tube into the supplied fitting directly in the process piping as shown below. Optional mounting brackets that attach to the instrument housing are also available.

WARNING: DO NOT CINCH DOWN OR TIGHTEN FERRULES OUTSIDE OF THE FERRULE REGION OF THE LONG TUBE OR YOU WILL RISK PERMANENTLY DAMAGING THE LONG TUBE AND SENSOR ASSEMBLY WITHIN.

1.0inches is the maximum distance that the fitting mount can be from the end plate of the Model 2230R unit. This is referred to as the Ferrule Region of the long tube which has a wall thickness of 0.065 in. Any distance exceeding 1.0 in. will be in the Sensor Assembly Region where the tube wall thickness is only 0.038 in. Cinching down a fitting outside the Ferrule Region and in the Sensor Assembly Region may result in permanent damage to the long tube and the sensor assembly within.

Page 9

Page 9
Image 9
Teledyne 2230R instruction manual Operation Unit Location, Mounting