MODEL 2230R, Process Hydrogen Analyzer

21

22

23

Settle time: 30 min

Type “ y”

(Y/N)?

 

 

 

Apply 30.000%H2:

Type “ y”

Ready (Y/N)?

 

 

Streaming data…

Taking Average…

Calibration Gas #2 finished.

cap=0.00000

Calibration complete.

res=x.xxxxx

 

4)Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of Calibration, conduct the Verification sequence as described in the previous section to verify that the calibration process was executed correctly.