MODEL 2230R, Process Hydrogen Analyzer
21
22 |
23
Settle time: 30 min | Type “ y” |
(Y/N)? |
|
|
|
Apply 30.000%H2: | Type “ y” |
Ready (Y/N)? |
|
| Streaming data… |
Taking Average… | Calibration Gas #2 finished. |
cap=0.00000 | Calibration complete. |
res=x.xxxxx |
|
4)Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of Calibration, conduct the Verification sequence as described in the previous section to verify that the calibration process was executed correctly.