MODEL 2230R, Process Hydrogen Analyzer

21

22

23

Settle time: 30 min

Type “ y”

(Y/N)?

 

 

 

Apply 30.000%H2:

Type “ y”

Ready (Y/N)?

 

 

Streaming data…

Taking Average…

Calibration Gas #2 finished.

cap=0.00000

Calibration complete.

res=x.xxxxx

 

4)Per TELEDYNE ANALYTICAL INSTRUMENTS’s recommendation, upon completion of Calibration, conduct the Verification sequence as described in the previous section to verify that the calibration process was executed correctly.

Page 26

Page 26
Image 26
Teledyne instruction manual Model 2230R, Process Hydrogen Analyzer