FIELD CALIBRATION STEPS

MODEL 2230R, Process Hydrogen Analyzer

Step

1

2

3

4

5

6

7

8

9

10

11

12

13

 

FIELD CALIBRATION STEPS

 

 

 

 

Display

 

User response

H2scan:

 

 

Type “ f” to run field calibration

 

 

 

 

Ready to Calibrate

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Pressure is 0.9690

 

 

Type “ y”

atm (Y/N)?

 

 

Please note that the firmware has remembered the

 

 

 

atmospheric pressure from the preceding Verification

 

 

 

procedure. Type “n” if a change is required and enter the

 

 

 

correct pressure.

 

 

 

 

Save as default

 

 

Type “ n”

(Y/N)?

 

 

 

Gas1 is 0%H2

 

 

Type “ n”

(Y/N)?

 

 

 

Gas1 for res (Y/N)?

 

 

Type “ y”

Cal Gas: 2.000%H2

 

 

Type “ n”

(Y/N)?

 

 

 

 

 

 

 

Enter gas:

 

 

Type “ 1” for gas #1, 1% hydrogen by volume

Cal Gas: 1.000%H2

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Settle time: 3 min

 

 

Type “ n”

(Y/N)? n

 

 

 

Enter time:

 

 

Type “ 30” for a 30 minute duration for gas #1

 

 

 

 

Settle time: 30 min

 

 

Type “ y”

(Y/N)?

 

 

 

 

 

 

 

Apply 1.000%H2:

 

 

Type “ y”

Ready (Y/N)? y

 

 

 

 

 

 

 

14

15

16

17

18

19

20

 

Streaming data…

Taking Average…

Calibration Gas #1 finished

cap=0.00000

 

res=x.xxxxx

 

Gas2 for res (Y/N)?

Type “ y”

Cal Gas:

Type “ n”

100.000%H2 (Y/N)?

 

 

 

Enter gas:

Type “ 30” for gas #2, 30% hydrogen by volume

 

 

Cal Gas: 30.000%H2

Type “ y”

(Y/N)?

 

Settle time: 1 min

Type “ n”

(Y/N)?

 

 

 

Enter time:

Type “ 30” for a 30 minute duration for gas #2