Teledyne 2230R Voltage Analog Output Range, 4.3SERIAL COMMUNICATION, Error, Serial, Software

Models: 2230R

1 26
Download 26 pages 44.62 Kb
Page 14
Image 14
Voltage Analog Output Range

MODEL 2230R, Process Hydrogen Analyzer

The user can change to another current range of the analog output in the field. Please refer to the “I” command in Section 4.3 SERIAL COMMUNICATION.

Below is the table for standard analog output voltage ranges:

Voltage Analog Output Range

0 V to 5 V

1 V to 5 V

0 V to 4 V

0.5 V to 4.5 V

Power-On Self Diagnostic

Error

 

 

0 V

5 V

 

 

0.5 V

0 V

 

 

5 V

4.5 V

0 V

5 V

The user can change to another voltage range of the analog output in the field. Please refer to the “V” command in Section 4.3 SERIAL COMMUNICATION.

WARNING: THE USER CANNOT CHANGE FROM A CURRENT RANGE TO A VOLTAGE RANGE OR FROM A VOLTAGE RANGE TO A CURRENT RANGE OUT IN THE FIELD. THIS REQUIRES A FACTORY MODIFICATION.

4.3SERIAL COMMUNICATION

The user can monitor output and interface with the unit to perform calibration or adjust user settings via the serial communication connector. The serial communication is accomplished via an RS232 (optional RS422) interface.

 

Serial

Any serial port two-way communications software such as terminal

 

 

Communications

emulators (HyperTerminal, Telnet, etc.) and purpose-built software

 

 

(using LabView, Visual Basic, C++, etc.) can be used to establish serial

 

 

Software

 

 

communication with the unit.

 

 

 

 

RS232 (RS422 optional)

 

 

Format and

19200 Baud

 

 

Settings

8 bit data

 

 

1 stop bit

 

 

 

 

 

 

 

 

No parity

 

 

 

 

Xon/Xoff

 

 

 

 

Streaming data is presented in column format. Once serial

 

 

 

 

communication is established and the unit is operating in normal

 

 

Data Display

 

mode, data will be displayed in the user specified format. The display

 

 

 

output options are configured via a serial command as described in the

 

 

 

 

following sections (refer to the SERIAL COMMUNICATION COMMANDS,

 

 

 

 

FORMAT <fmt>, and OPTIONS <opt> sections). Columnated data

 

 

 

 

available are as follows:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 14

 

Page 14
Image 14
Teledyne 2230R Voltage Analog Output Range, 4.3SERIAL COMMUNICATION, Error, Serial, Communications, Software, Format and