Include resistor reading

MODEL 2230R, Process Hydrogen Analyzer

Include resistor reading

0

0

0

0

 

0

1

0

0

Include overall hydrogen reading

0

0

0

0

 

0

0

1

0

Include peak hydrogen reading

0

0

0

0

 

0

0

0

1

 

 

 

The Options <opt> string is a two character hexadecimal representation

Options <opt>

of an 8 bit value derived from the following table. The user determines

which data is needed and selects that bit value. Once all selections are

 

made the values are summed bitwise and then converted to a two

 

place hexadecimal value. To aid in the conversion, a 4 bit-to-

 

Hexadecimal conversion table follows in EXAMPLE 3.

 

 

 

 

Options: Bit Value Identifiers

Serial output Options <opt> parameter: select the desired status messages from this list which will appear in the MESSAGES column, add the bit value for each column bitwise, and convert into two hexadecimal characters using the 4 bit-to-Hexadecimal table in EXAMPLE 3.

Description

 

 

 

Bit Value

 

 

 

 

 

 

 

 

 

Calculation Errors

0

0

0

0

0

1

0

0

Heater State

0

0

0

0

0

0

1

0

 

 

 

 

 

 

 

 

 

Example 3:

The user wishes to implement the “G” command (“Go”command, refer to LEVEL 1 COMMANDS table) to have the following serial output columns reported from the monitor: Time Stamp, Capacitor Reading, Overall Hydrogen Reading, the Peak Hydrogen Reading, Calculation Errors, and the Heater state.

From the FORMAT <fmt> table above, you identify your desired columns with its corresponding bit value:

<fmt> Descriptions

<fmt> Bit Value

Time Stamp

1000 0000

Capacitor Reading

0000 1000

Overall Hydrogen

0000 0010

Reading

 

Peak Hydrogen

0000 0001

Reading

 

<fmt> 4 Bit Value

1000 1011

Combination:

 

4 bit value

Hexadecimal

 

Character

0000

0

0001

1

0010

2

0011

3

0100

4

0101

5

0110

6

0111

7

1000

8

1001

9

1010

A

1011

B

1100

C

1101

D

1110

E

1111

F