3.0 Startup

3.0 Startup

Perform the following steps when fist installing the M460H ozone monitor or when bringing the monitor back into service.

1.Turn on power to the monitor. The display will briefly display the “API” logo followed by the software version. The display will then begin showing ozone concentration.

2.Establish a flow of zero gas to the monitor. Zero gas can be oxygen or clean, dry air. If monitor cannot be disconnected from an ozone generator, then establish flow through the generator and monitor with the generator turned off. Flow rate through monitor should be between 0.5 – 2.0 LPM (Liters per minute.)

3.Let monitor warm up and purge for a minimum of 5 minutes. Zero the monitor by simultaneously pressing the Zero buttons on the front panel. The display should momentarily show dashes (‘----‘) after which the display should go quickly to zero.

4.Check Status LED’s on front panel; Sensor OK led should be ON, all other LED’s should be OFF. If the Status LED’s are not in this state, refer to Chapter 7 for troubleshooting information.

5.Observe the monitor for several more minutes at zero to ensure that it is stable.

6.The monitor is now ready to measure ozone.