Teledyne 460H Setaddr, addressSETADDRnewaddressCR, Syntax, Description, Data Parameters Sent

Models: 460H

1 38
Download 38 pages 41.04 Kb
Page 27
Image 27

5.4.5 SETADDR

Syntax

<address>SETADDR:<new_address><CR>

Description

Changes the communications address to a new value.

Data Parameters Sent

new_address is the new address for the monitor; Range for new_address is 1-9.

Response

<address>:<success_flag>#<checksum><CR>

Example

Command:

1SETADDR:2<CR> - Change address from 1 to 2

Response:

1:OK#261<CR> - Change Address Successful 1:FAIL#391<CR> - Change Address Failed

Notes

Note that monitor response is from previous address, but any further commands must be at new address. In the example above, after sending back the OK response, the monitor will no longer respond to commands with address 1.