4.2 Auto-Zero Operation

4.2 Auto-Zero Operation

The Auto-Zero feature allows the monitor to operate unattended for extended periods of time. This feature automatically performs a Zero Calibration at specific intervals (default interval is 12hrs.) The monitor does this by opening a solenoid valve to allow externally supplied Zero Gas to purge the monitor and then automatically performs a Zero Calibration.

4.2.1 Auto-Zero Installation

In order for this feature to function properly, external, pressurized zero gas must be supplied to the monitor as detailed in Section 2.2.

4.2.2 Auto-Zero Function

When the Auto-Zero function is triggered, the monitor performs the steps outlined in Table 4-1 below.

Table 4-1 Auto-Zero Function

Step

Function

Display Reads

1

Switch Auto-Zero valve on

N/A

2

Purge cell for 20 seconds

Holds and Flashes Concentration Value

3

Perform Zero Calibration

Dashes: ‘----‘

4

Switch Auto-Zero valve off

Dashes: ‘----‘

5

Hold-Off for 3 sec to return

Dashes: ‘----‘

 

Measurement Gas to cell

 

6

Return to Measurement Mode

Normal Concentration Display

4.2.3 Auto Zero Error

When performing an Auto-Zero Calibration, the monitor checks to make sure the ozone concentration is near zero before calibration. If the ozone concentration is above 0.5wt% or 5.0 g/Nm3, the monitor will not perform the calibration and will display a “ZERO ERR” message on the display. The monitor will also turn on the Auto-Zero Error status signal (See Section 2.4.2 Digital Status Outputs.) This Status Signal will remain on until another Auto-Zero Calibration is successfully performed or a Manual Zero Calibration is performed.

If an Auto-Zero error is detected, a manual Zero Calibration should be performed (See Section 4.1.3 Zero Calibration.)

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 20

Page 20
Image 20
Teledyne 460H Auto-Zero Operation, Auto-Zero Installation, Auto Zero Error, 1 Auto-Zero Function, Step, Display Reads