Teledyne 460H Tlist, Syntax, Description, Data Parameters Sent, Response, Example

Models: 460H

1 38
Download 38 pages 41.04 Kb
Page 29
Image 29

5.4.7 TLIST

Syntax

<address>TLIST<CR>

Description

Returns a verbose list of the current measured parameters with labels for each parameter: O3 Concentration, Cell Pressure(psia,) Cell Temperature (K,) Lamp Temperature (K,) Measure Detector (mV,) Calibrated Reference Detector (mV,) Reference Detector (mV)

Data Parameters Sent

None

Response

Test Parameter List (See Below)

Example

Command:

1TLIST<CR>

Response:

O3 = 0.0226168

Press = 14.7753

Cell Temp = 300.7116

Lamp Temp = 324.7965

Ref = 2881.52

Meas = 2941.092

Raw Ref = 4412.646

Notes

Note that the concentration value returned shows more digits after the decimal than the front panel display. Concentration values should only be considered valid to 4 significant digits.

Also, note that no checksum is sent in the response to the TLIST command.

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 29

Page 29
Image 29
Teledyne 460H instruction manual Tlist, Syntax, Description, Data Parameters Sent, Response, Example