7.6 Status Output Summary Table

7.6 Status Output Summary Table

Table 7-2 below is a logic truth table summarizing the recommended actions based on the states of the four status outputs. A ‘1’ indicates the output is ON, a ‘0’ indicates the output is OFF, and ‘X’ indicates the output is in either state.

Table 7-2: Status Output Truth Table

Sensor

Invalid

Lamp

Cell

Actions

OK

Reading

Low

Dirty

 

1

0

0

0

Normal operation, no action

 

 

 

 

required

0

X

X

X

Service required

1

1

X

X

Check Pressure > 45 psia

 

 

 

 

Verify that concentration has

 

 

 

 

not exceeded full scale range

 

 

 

 

of sensor.

 

 

 

 

Calibrate at Zero.

1

X

1

X

Lamp adjustment useful,

 

 

 

 

though not required

1

X

X

1

Calibrate at zero

 

 

 

 

Clean Cell

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 36

Page 36
Image 36
Teledyne 460H Status Output Summary Table, 2 Status Output Truth Table, Sensor, Invalid, Lamp, Cell, Actions, Reading