7.0 Sensor and System Troubleshooting
This chapter gives guidelines for diagnosing system and sensor malfunctions using the five digital Status Outputs provided by the M460H. All troubleshooting should be done after the M460H has been turned on and allowed to warm up for at least 15 minutes.
7.1 Status Output Summary
Table
Table 7-1: Digital Status Outputs
Output | Name | On State | Off state |
# |
|
|
|
1 | Sensor O.K. | Normal State | Reference or |
|
|
| Measure > 4995mV; |
|
|
| Reference < 1000mV |
2 | Invalid Reading | Pressure > 45 psia, | Normal State |
|
| Negative Ozone |
|
|
| Concentration, |
|
|
| Concentration Over- |
|
|
| Range |
|
3 | Lamp Low | Reference | Normal State |
|
| Detector<2500mV |
|
4 | Cell Dirty | Measure/Reference | Normal State |
|
| ratio < 0.5 (zero gas) |
|
5 | Normal State |
7.2 Sensor OK
The normal state for the Sensor OK output in ON. During the
If the Sensor O.K. output turns off AND the Lamp Low output is on, this indicates that the lamp intensity has below the minimum level required for proper operation.
If the Sensor O.K. output turns off and the Lamp Low output is also off, then one of the analog voltages in the sensor has exceeded the range of the internal A/D converter. Adjustment by qualified service personnel is required.