Teledyne 460H Command Summary, Examples of valid Commands with no data, Description

Models: 460H

1 38
Download 38 pages 41.04 Kb
Page 22
Image 22
Examples of valid Commands with no data:

Examples of valid Commands with no data:

Checksum Included:

1O3#179<CR> No Checksum:

1O3<CR>

Examples of valid Commands with data:

Checksum Included:

1VSET:1,20#620<CR> No Checksum:

1VSET:1,20<CR>

5.3 Command Summary

Table 5-2 below lists the commands available and a summary of their function.

Table 5-2 Serial Port Command Summary

 

Command

 

 

Description

 

 

 

 

 

CAUTO

 

Perform an Auto-Zero Calibration (if option is installed)

CZERO

 

Perform a manual zero calibration (zero calibration is

 

 

 

computed based on current O3 reading in cell)

DACSTEP

 

Analog Output Test Mode, Step Function

O3

 

Returns current O3 concentration

SETADDR

 

Set communications address

TDUMP

 

Returns a compact list of current measurement parameters

TLIST

 

Returns all current measurement parameters

VGET

 

Get Single VAR value

VLIST

 

Lists all internal VARS

VSET

 

Sets value of internal VAR

P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 22

Page 22
Image 22
Teledyne 460H Command Summary, Examples of valid Commands with no data, Examples of valid Commands with data, Description