Examples of valid Commands with no data:
Checksum Included:
1O3#179<CR> No Checksum:
1O3<CR>
Examples of valid Commands with data:
Checksum Included:
1VSET:1,20#620<CR> No Checksum:
1VSET:1,20<CR>
5.3 Command Summary
Table
Table 5-2 Serial Port Command Summary
| Command |
|
| Description |
|
|
|
|
| ||
CAUTO |
| Perform an | |||
CZERO |
| Perform a manual zero calibration (zero calibration is | |||
|
|
| computed based on current O3 reading in cell) | ||
DACSTEP |
| Analog Output Test Mode, Step Function | |||
O3 |
| Returns current O3 concentration | |||
SETADDR |
| Set communications address | |||
TDUMP |
| Returns a compact list of current measurement parameters | |||
TLIST |
| Returns all current measurement parameters | |||
VGET |
| Get Single VAR value | |||
VLIST |
| Lists all internal VARS | |||
VSET |
| Sets value of internal VAR |
P/N 03662D Teledyne API Model 460H O3 Monitor Operator Manual - Page 22