Infrared Gas Analyzer Installation
Teledyne Analytical Instruments 27
Table 2-1: Correspondence of Measured Components and Measuring
Units
Measuring components Measuring unit 1 Measuring unit 2
1-component meter for NO,
SO2, CO2, CO and CH4 Each component None
2-component meter for
NO/SO2 and CO2
/
CO NO/SO2
CO2
/
CO None
2-component meter for
NO/CO NO CO
3-component meter for
NO/SO2/CO NO/SO2 CO
4-component meter for
NO/SO2/CO2/CO NO/SO2 CO2/CO
2.5.2 External Piping Diagram
There are several ways to bring sample gas to the analyzer
depending on the number of inlet/outlet gas connections on your
instrument Recommended piping diagrams are shown in Figures 2-5, 2-
6 and 2-7. Figure 2-5 is a schematic for an instrument with a single pair
of inlet/outlet gas connections. Figures 2-6 and 2-7 are used when there
are two pair of inlet/outlet connections. Note that a NO2/NO converter
is used when NO measurement is used for NOx analysis.
Figure 2-5: External Piping with Single Inlet and Outlet