Teledyne 7600 operating instructions Flowrate, Preparation of Calibration Gas

Models: 7600

1 126
Download 126 pages 58.79 Kb
Page 29
Image 29

Infrared Gas Analyzer

Installation

 

 

 

The dew point of the sample gas must be lower than the ambient temperature to avoid condensation in the analyzer. If vapor is contained in the sampling gas, the dew point should be lowered to 0°C by using a dehumidifier.

If SO2 mist is contained in the sample gas, use a mist filter or cooler to remove the SO2 mist. Other mists should be removed by using an appropriate mist filter or cooler.

Corrosive gases such as Cl2, F2 and HCl, if contained in the sample gas in considerable amounts, will shorten the life of the instrument.

Sample gas temperature should be within 0 to 50°C. Hot gas should not be fed directly into the instrument.

2.5.4Flowrate

Use a flowrate of 0.5L/min ±0.2L/min. The sample system should be designed to avoid any flow fluctuation during measurement.

Use a flowmeter to observe the flow reading as shown in the external piping diagrams of Figures 2.5, 2.6 and 2.7.

2.5.5 Preparation of Calibration Gas

Routine calibration using calibration gases is required for optimizing the performance of this instrument. Once a week is a suggested calibration frequency. Table 2-2 indicates the zero and span gas required for calibration.

Table 2-2: Zero and Span Gas

 

Analyzer without

Analyzer with built-in

Analyzer with external

 

O2

O2

zirconia O2 sensor

Zero gas

N2 gas

N2 gas

Dry air

 

 

 

 

Span gas

Gas concentration

Gas concentration of

Gas concentration of 90%

other than for

of 90% or more of

90% or more of full

or more of full scale

O2

full scale

scale

 

t

 

Gas concentration of

1 to 2% O2

Span gas for

NA

O2 analysis

90% or more of full

 

 

 

scale

 

Teledyne Analytical Instruments

29

Page 29
Image 29
Teledyne 7600 operating instructions Flowrate, Preparation of Calibration Gas