Infrared Gas Analyzer | Installation | |
|
|
|
•The dew point of the sample gas must be lower than the ambient temperature to avoid condensation in the analyzer. If vapor is contained in the sampling gas, the dew point should be lowered to 0°C by using a dehumidifier.
•If SO2 mist is contained in the sample gas, use a mist filter or cooler to remove the SO2 mist. Other mists should be removed by using an appropriate mist filter or cooler.
•Corrosive gases such as Cl2, F2 and HCl, if contained in the sample gas in considerable amounts, will shorten the life of the instrument.
•Sample gas temperature should be within 0 to 50°C. Hot gas should not be fed directly into the instrument.
2.5.4Flowrate
Use a flowrate of 0.5L/min ±0.2L/min. The sample system should be designed to avoid any flow fluctuation during measurement.
Use a flowmeter to observe the flow reading as shown in the external piping diagrams of Figures 2.5, 2.6 and 2.7.
2.5.5 Preparation of Calibration Gas
Routine calibration using calibration gases is required for optimizing the performance of this instrument. Once a week is a suggested calibration frequency. Table
Table 2-2: Zero and Span Gas
| Analyzer without | Analyzer with | Analyzer with external |
| O2 | O2 | zirconia O2 sensor |
Zero gas | N2 gas | N2 gas | Dry air |
|
|
|
|
Span gas | Gas concentration | Gas concentration of | Gas concentration of 90% |
other than for | of 90% or more of | 90% or more of full | or more of full scale |
O2 | full scale | scale |
|
t |
| Gas concentration of | 1 to 2% O2 |
Span gas for | NA | ||
O2 analysis | 90% or more of full |
| |
|
| scale |
|
Teledyne Analytical Instruments | 29 |