Operation Model 7600
Teledyne Analytical Instruments 66
3. In the “Setting of Auto Zero Calibration” screen that
appears, point the cursor to any item you want to set by
pressing the UP or DOWN key. Press the ENT key.
4. In the “Auto Zero Calibration Parameter Setting” screen
that appears, perform the value entry or the setting. For
the value entry or setting change, use the UP or DOWN
key. To change the setting, use the SIDE key to move the
cursor to the right.
5. After changing the setting, press the ENT key to accept
the change. Auto zero calibration will now be carried out
using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
The contacts for the auto zero calibration are closed during auto
zero calibration.
Auto Zero Calibration Example
Start Time SUN 12:00
Cycle 12 hour
Flow Time 300 sec
ON/OFF ON
Figure 3-8 shows the auto zero calibration with the above settings.