Operation Model 7600
Teledyne Analytical Instruments 58
To close the “Hysteresis Setting” or cancel the mode midway, press
the ESC key. A previous screen will return.
3.5.4 Setting of Auto Calibration
3.5.4.1 AUTO CALIBRATION
Auto calibration is automatically carried out when zero and span
calibrations are set.
In this menu you can set when and how frequent the auto
calibration is performed. You also enable or disable the auto calibration
feature here.
Description of setting items:
Start Time: Setting at the first calibration (day of the week, hour,
minute)
Cycle: A period between the start time of one calibration and
another (unit: hour/day)
Flow Time: The time required for replacement by calibration gas.
Time required for replacement of sample gas after the
calibration is completed (Set by calibration gas.)
ON/OFF: Enable/disable the auto calibration feature.
Note: Before changing the settings of auto calibration, set the
ON/OFF to OFF.
To change the settings:
3. From the measurement mode, press the MODE key to
display the user mode.
4. Point the cursor to “Setting of Auto Calibration” by
pressing the UP or DOWN key and press the ENT key.