To close the “Hysteresis Setting” or cancel the mode midway, press the ESC key. A previous screen will return.
3.5.4 Setting of Auto Calibration
3.5.4.1AUTO CALIBRATION
Auto calibration is automatically carried out when zero and span calibrations are set.
In this menu you can set when and how frequent the auto calibration is performed. You also enable or disable the auto calibration feature here.
Description of setting items:
Start Time: Setting at the first calibration (day of the week, hour, minute)
Cycle: | A period between the start time of one calibration and |
| another (unit: hour/day) |
Flow Time: The time required for replacement by calibration gas. Time required for replacement of sample gas after the calibration is completed (Set by calibration gas.)
ON/OFF: Enable/disable the auto calibration feature.
Note: Before changing the settings of auto calibration, set the ON/OFF to OFF.
To change the settings:
3.From the measurement mode, press the MODE key to display the user mode.
4.Point the cursor to “Setting of Auto Calibration” by pressing the UP or DOWN key and press the ENT key.