Infrared Gas Analyzer

Maintenance

 

 

 

Error

Error Contents

Probable Cause

Display

 

 

Error No.5

Amount of zero calibration

• Zero is deflected much due to dirty

 

(indication value) is over 50% of

cell.

 

full scale.

• Detector is faulty.

 

 

 

 

• Optical balance is out of adjustment.

Error No.6

Span calibration is not within the

• Span gas is not supplied.

 

allowable range.

• Calibrated concentration setting does

 

 

Error No.7

Amount of span calibration

not match cylinder concentration.

 

 

(difference between indication

• Zero calibration was performed

 

value and calibrated

incorrectly.

 

concentration) is over 50% of full

• Dirty cell.

 

scale.

• Detector sensitivity has deteriorated.

 

 

Error No.8

Measured values fluctuate too

• Calibration gas is not supplied.

 

much during zero and span

• Time for flowing calibration gas is too

 

calibration.

 

short.

 

 

Error No.9

Calibration is abnormal during

• No. 4 to No. 8 Error occurred during

 

auto calibration.

 

Error No.10

Output cable connection is

• Wiring is detached between analyzer

 

improper.

and interface module.

 

 

• Wiring is disconnected between

 

 

analyzer and interface module.

When errors No. 1 and No. 10 occur, the analyzing block error contact is closed. When errors No. 4 to No. 9 occur, calibration error contact is closed.

When an error occurs, the error number message will appear at the offending channel. The screen below shows that error No. 9 has occurred on Channel 1.