Installation Model 7600
Teledyne Analytical Instruments 32
2.6 Electrical Connections
2.6.1 Power Inlet
The power inlet for the Model 7600 is located on the rear panel as
shown in Figure 2-9. The proper 3-prong power cord has been supplied
with your instrument. Connect the power cable to the power inlet.
Figure 2-9: Model 7600 Electrical Connections
The Model 7600 requires 100-240 VAC 50/60 Hz power.
Avoid installing this instrument near an electrical unit that
generates electrical noise such as a high frequency furnace or electric
welder. If using the instrument near such a noise-generating unit is
unavoidable, use a separate power line to avoid noise.
Mount a noise suppressor such as varister or spark killer as shown
in Figure 2-10 to the noise generating unit when noise is generated from
relays or solenoid valves. The suppressor must be mounted near the
noise generating source, or it will have no effect.