Operation

Model 7600

 

 

 

5.On the parameter hold screen that appears, move the cursor next to the channel you want to make the setting by pressing the UP or DOWN key, and then press ENT.

6.The value is highlighted, indicating that the value can be changed. Change the numerical value by pressing the UP or DOWN key, and then move the cursor to the right by pressing the SIDE key.

7.After the value is changed, press the ENT key to save.

8.Press the ESC key to return to the parameter setting screen.

The setting is expressed in % against the range for both ranges. For example, when 0 to 1000 ppm is selected as the range, and 10% FS is selected as hold setting, the output equivalent of 100 ppm is output and held irrespective of the measurement value at that time.

Description of setting:

Instantaneous value display of the measurement cannot be held. Only the output can be held.

If set value is selected for hold, instantaneous O2 correction value is calculated and held based on the set value.

Range identification contact output cannot be switched even if the range is switched during the hold.

3.5.7.2AVERAGE VALUE RESET:

This mode is used to clear all average values O2 correction average and O2 average, and restarts averaging. All average values are reset at the same time. The indication value and output value is 0 ppm, vol % etc. at the time of the reset input (Refer to the average period). To