18.1 Trim
When the pump is under remote control, it tracks an analogue signal from the user’s remote control system to the i/p terminal of the Analogue 1 and Analogue 2 con- nectors at the rear of the pump within the ranges
Select Trim from the Setup menu or the Analogue setup menu using the UP or DOWN keys and press ENTER to confirm your choice.
Apply the low process analogue signal to the i/p terminal of the Analogue 1 connector as instructed in the display. See 18.2 Analogue. While the signal is being applied, press ENTER to record the signal as a calibration point.
Apply the maximum process control signal. While the signal is being applied, press ENTER to record the signal as a calibration point.
Apply 50% of the maximum process control signal. While the signal is being applied, press ENTER to record the signal as a calibration point.
If a mistake is made, press STOP at any point in the sequence, and the pump displays the previous screen.
36 |