29.2 520R and 520R2 tube loading
520R continuous tubing pumpheads are
Mark a 225mm (8 7⁄8 in) length onto the section of the tubing which is to be located into the pumphead. Caution: Placing more than 225mm (8 7⁄8 in) of tube inside the pumphead may result in reduced tube life.
Open the lower
Disengage the rotor clutch by fully depressing the yellow clutch button on the side of the rotor hub and turning the hub a few degrees while the clutch button is still depressed. The rotor can now rotate independently of the gearbox and motor for one full revolution. If the clutch
Feed the tubing around the pumphead track, turning the rotor as necessary. Make sure the tubing is not twisted or pinched between the guide rollers and the track. Ensure that the second 225mm (8 7⁄8 in) mark is adjacent to the inner edge of the upper tube clamp.
Open the upper
84 |