ResMed AutoSet CS 2 user manual Troubleshooting, Problem Possible cause Solution

Page 43

TROUBLESHOOTING

If there is a problem, try the following suggestions. If the problem cannot be solved, contact your equipment supplier or ResMed. Do not attempt to open the flow generator.

Problem

Possible cause

Solution

 

 

 

Insufficient air delivered from

Air filter is dirty.

Replace air filter.

flow generator

 

 

 

Air tubing is kinked or

Straighten or replace tubing.

 

punctured.

 

 

 

 

Flow generator operating

Water may have entered the

Remove the water from the

erratically, ie, switching on and

air tubing.

air tubing.

off, pressure varying at random

 

 

 

 

 

Pressure rises inappropriately

You are talking, coughing,

Avoid talking with the mask

 

voluntarily breath-holding (eg

on, and breathe as normally

 

while rolling over in bed), or

as possible.

 

intentionally breathing in an

 

 

unusual manner.

 

 

Severe mask leak or mouth

Correct your leak condition.

 

leak (greater than 30 L/min).

 

 

 

 

Alarm stays on continuously

Blockage in air circuit.

Ensure there is no barrier to

 

 

the passage of air.

 

Internal malfunction.

Return the device to your

 

 

equipment supplier for

 

 

service.

 

 

 

No LCD display

Power not connected or

Ensure the power cable is

 

switch at back is not on.

connected, and that the

 

 

switch at the back of the unit

 

 

is in the “on” position.

 

 

 

FAULT message displayed on

Internal malfunction.

Return the device to your

LCD

 

equipment supplier for

 

 

service.

ENGLISH

System Error 7 message

Device power-on has been

 

interrupted by breathing into

 

the mask during warm up.

Turn the flow generator off and on again. Do not place mask on face during warm up (30 seconds). Select LEARN CIRCUIT (page 26). If the error is not cleared, contact your equipment supplier for service.

Internal malfunction.

Return the device to your

 

equipment supplier for

 

service.

 

 

TROUBLESHOOTING

39

Image 43
Contents AutoSet CS EUR AutoSet CS Page Autoset CS Page Table of Contents Cleaning and Maintenance Medical Information IntroductionAutoset CS2 USER/OWNER ResponsibilityPage Quick Setup Guide Page Autoset CS2 Components Autoset CS2 SystemConnect this end Assembling the Autoset CS2 System Connect the power cordConnect air tubing Attach the mask to the air tubing HumidifiersConnecting a Humidifier Humidaire Perform the Air Circuit Learn function see ANTI-BACTERIAL Filter ReslinkInserting anti-bacterial filters Using DC Power to RUN the Autoset CS2Oxygen tube attached to Mirage FFMS2 and Ultra Mirage masks Using Supplemental OxygenPage Stopping Treatment Using the Autoset CS2Standby Mode Starting TreatmentUP/DOWN KEY Control Panel FunctionsLCD Screen START/STOP KEYRight KEY RED Adjusting SettingsAlarm Leds and Alarm Mute Left KEY GreenExample You wish to change the sound level for the alarmsNavigating the Patient menu MenusLearn Circuit MaskLeak Alert SmartstopSound Level Smart Data Mask FITAverage Pressure UsageApnea Treatment ScreensPage If a LOW Pressure Alarm Sounds AlarmsTriggered by Will Clear When If a Power Fail Alarm SoundsIf a High Pressure Alarm Sounds If a High Leak Alarm SoundsIf a Flow Blocked Alarm Sounds If AN Insufficient Pressure Support Alarm SoundsPage Daily Cleaning and MaintenanceWeekly AIR Filter PeriodicallyFitting the Filter ServicingPage Problem Possible cause Solution TroubleshootingPage Glossary Page Technical Specifications Leak exceeds 30 L/min for more than 20 seconds Max Pressure Environmental information English Next Recommended separation distance Separation distance according to frequency of transmitter Product Warranty Period Limited WarrantyPage Index Page Manuel Utilisateur Page Sommaire Nettoyage ET Entretien Informations Medicales Responsabilite DE L’UTILISATEUR/DU ProprietaireAvertissements PrecautionsGuide D’INSTALLATION Rapide Page Composants DE L’AUTOSET CS2 Autoset CS2Raccordez cette extrémité Montage DE L’AUTOSET CS2 AvertissementRaccordez le circuit respiratoire Montage de l’AutoSet CS2 HumidificateursVoir « Début du traitement » De l’AutoSet CS2 Avertissement Filtre Antibacterien Mise en place des filtres antibactériens Utilisation D’UNE Alimentation CC Avec L’AUTOSET CS2Utilisation D’OXYGENE Supplementaire AvertissementsPage Arret DU Traitement Mode D’ATTENTEUtilisation DE L’AUTOSET CS2 Debut DU TraitementTouche MARCHE/ARRET Touche DE Deplacement Vers LE HAUT/BASFonctions DU Panneau DE Controle Ecran LCDTouche DE Droite Rouge Temoins D’ALARME ET Desactivation DE L’ALARMEModification DES Reglages Touche DE Gauche VerteExemple Vous souhaitez modifier le niveau du son pour les alarmesNavigation dans le menu Patient IDENT. Circuit MasqueAI Faible Alarme FuitePression Moyenne Ajustement DU MasqueUtilisation RETRO-ECL Ecrans DE TraitementLangue Remarquecibl nest pas un paramètre utilisé en mode Cpap Declenchee Lorsque LES Alarmes’ARRETE Lorsque SI UNE Alarme DE Pression Basse SE DeclencheSI UNE Alarme DE Panne DE Courant SE Declenche SI UNE Alarme DE Pression Elevee SE DeclencheSI L’ALARME D’AIDE Inspiratoire Insuffisante SE Declenche SI UNE Alarme DE Fuite Importante SE DeclencheSI L’ALARME DE Debit Bloque SE Declenche Page Quotidien Nettoyage ET EntretienHebdomadaire Démontage du circuit respiratoire pour le nettoyage PeriodiqueMise EN Place DU Filtre Filtre a AIREntretien Strategie DE Depannage Problème Cause possible SolutionPage Glossaire Page Caracteristiques Cm H2O, si cette valeur est inférieure Pression Max Informations concernant l’environnement 101 102 103 104 Produit Durée de la Garantie Garantie Limitee106 Index 108

AutoSet CS 2 specifications

ResMed AutoSet CS 2 is a state-of-the-art device designed for individuals suffering from complex sleep apnea and central sleep apnea. This advanced treatment solution leverages innovation and cutting-edge technology to provide optimal comfort and effective therapy during sleep, ultimately helping users achieve a better quality of life.

One of the standout features of the AutoSet CS 2 is its sophisticated auto-adjusting pressure capabilities. The device automatically detects and responds to the patient's breathing patterns throughout the night, adjusting the pressure levels in real time to ensure that airway obstructions are effectively managed. This technology helps prevent both hypopneas and apneas, providing a seamless therapy experience.

The device is equipped with ResMed's proprietary Easy-Breathe technology, which allows for smooth pressure changes that mimic the user's natural breathing cycle. This feature enhances user comfort, making it easier for individuals to adapt to CPAP therapy and continue using the device consistently over time.

Another remarkable characteristic of the AutoSet CS 2 is its integrated wireless connectivity. This connectivity facilitates remote monitoring and data sharing with healthcare professionals, enabling more effective management of treatment progress. By analyzing sleep data, providers can make informed adjustments to therapy, ensuring that the patient receives the best possible care.

The AutoSet CS 2 also includes a user-friendly interface with a high-resolution color display, making it easy for users to navigate menus and access vital information about their sleep patterns and therapy effectiveness. With its compact design, the device is portable and can easily be used at home or while traveling.

In terms of comfort, the AutoSet CS 2 features a quiet operation, enabling a peaceful sleep environment. It is compatible with various masks, providing flexibility for users to find the best fit for their personal comfort preferences.

Overall, the ResMed AutoSet CS 2 is a leading choice for individuals with complex sleep apnea. With its combination of advanced features, user-centric design, and state-of-the-art technology, it empowers users to take control of their sleep and improve their overall health and well-being.