Carrier 500 T/R user manual Day start time, Day end time, Alarm menu

Page 14

3.1.7Day start time

A print out starts from the actual time and goes backwards until the time set in this parameter, unless the ‘Print time period’ is shorter. If the ‘Day start time’ is set to “00:00” only the ‘Print period time’ is applicable. This allows you to print back to a maximum of 24 hours.

3.1.8Day end time

This option is only used when printing a historical date. The print will start with the time set in this option and goes backwards until the required ‘Day start time’. The ‘Print time period’ is not applicable.

3.2Alarm menu

1.7Day start time (06:00)

↑ ↓ edit <-menu

1.8Day end time (18:00)

↑ ↓ edit <-menu

Up to four different alarm types can be allocated to up to four different compartments. The various alarms (and compartments) are only available if the supervisor has pre-set and configured them in the parameter menu.

By pressing the yellow button the alarm menu will open. With [↑], [↓] you can toggle between the pre-set compartments. Between ‘(..)’ the actual setting is displayed. By pressing [edit] you can determine an alarm for the relevant compartment.

With [<], [>] you select the desired alarm type (including alarm off). By pressing [accept] the chosen alarm will be activated.

2.1

Comp. 1

 

(off)

 

 

edit

<-menu

 

 

 

 

 

 

2.1

Comp.1

 

Frozen

 

 

<

>

accept

<- cncl

 

 

 

 

Note : In case of an alarm condition, an acoustic signal will be audible and on the display the corresponding temperature input will flash.

After an alarm is accepted with the yellow button the acoustic signal will stop but the alarm remains active until the temperature is again within the limits (for each compartment the desired alarm can be activated).

As soon as a protocol is selected in the recorder to communicate with the micro of an Advance Microprocessor, TM or R/TR unit, we have in Menu 2.1, the name COMP1 for compartment 1 set automatically to “on”.

The “off” choice can be selected by the user to disable the alarm. The Pre-set alarm groups are not available anymore at this level.

The same principle is applicable for compartment 2 and 3.

Page 14

62-61138-20 (02/03)

Image 14
Contents Installation USER’S Manual DataCOLD 500 R DataCOLD 500 TTable of Contents Yellow General descriptionBlue Green EditPositioning Connections InstallationPrinter Optional junction box MountingDataCOLD R-Version Optional Mounting kit CON4DataCOLD T-Version ConnectorsPower supply Is not Used for More than TEN 10 Days Configuration Alarm outputsWiring Display background lighting preferable for the R-VersionSupplementary installation Final testing Standard InstallationTemperature sensors Alarm signalUser menus Page Day end time Alarm menuDay start time Summer / wint time setting User settingsTime setting Date settingBuzzer volume setting Display back light settingKeyboard back light setting Buzzer frequency settingT1 Input On/off Menu 5 Temperature inputs settingsParameter menu Return airT3 Input On/off. Menu structure as T4 Input Menu 7 Compartment settings Menu 6 Digital inputs settingsDigital inputs setting On/off Compartment On/offMenu 9 Printer settings Menu Alarm settingsAlarm group On/off Delivery ticket settingsMenu 11 Communication settings Menu 10 General settingsTemperature units C 10.2 Sample rate Minutes11.2 COM2 RS-232 port setting Accessories and spare parts Part Number Designation62-61138-20 02/03 Enclosure a Enclosure B Replace paper roll Compartment settings Enclosure C Factory settingsTemperature inputs Factory setting Printer settingsEnclosure D Failure codes & Set correction factor procedure Enter PincodeEnclosure E Wiring diagrams Digital Inputs configuration example= GND = RXD = TXD = CTS