Enclosure D Failure codes & Set correction factor procedure

If the display shows one of the following values:

OC

(I)

the input is activated but no sensor is connected.

 

(ii)

the sensor has an open circuit (sensor or wire failure.)

SC

(I)

the sensor has a short (sensor or cable failure.)

Procedure to set a Correction factor

The DataCOLD 500 R/T recorders have an option to adjust the offset of the A/D converter with a correction factor. Because this offset is subject to aging it is recommended to repeat this procedure on a regular basis (i.e. at the annual reference check).

The correction factor is a fixed parameter used in the formula which converts the measured resistance into the corresponding temperature. This procedure is not applicable for calibrating individual sensors. Note that the effect on the measured and displayed Temperature is not linear.

Set the correction factor

Connect to Temperature input 1 a precision resistor of 1500 Ω (0,1%).

Enter the parameter menu by pressing the green button for 2 seconds. The display will show: "ENTER PINCODE". The factory setting of the pin code is 1111.

After entering the correct pin code you have access to the parameter menu’s. Select menu 10.6 and press [edit].

The recorder will now compare the measured value of the resistor with the ‘expected’ value of 1500 Ω.

The second line the recorder will display a recommended correction factor based on the measured value. This factor can be set on line three.

Parameter Settings

ENTER PINCODE

1

****

3

4

2

10.5Corr.factor Recommended: +2 Setting: +1

< > ACCEPT <-CNCL

After pressing ‘accept’ this correction factor will be applicable for all Temperature inputs. The correction factor is limited from -9 to +9.

Page 28

62-61138-20 (02/03)

Image 28
Contents Installation USER’S Manual DataCOLD 500 R DataCOLD 500 TTable of Contents General description BlueYellow Green EditPrinter InstallationPositioning Connections DataCOLD R-Version MountingOptional junction box Optional Mounting kit CON4Power supply ConnectorsDataCOLD T-Version Is not Used for More than TEN 10 Days Alarm outputs WiringConfiguration Display background lighting preferable for the R-VersionFinal testing Standard Installation Temperature sensorsSupplementary installation Alarm signalUser menus Page Day start time Alarm menuDay end time User settings Time settingSummer / wint time setting Date settingDisplay back light setting Keyboard back light settingBuzzer volume setting Buzzer frequency settingMenu 5 Temperature inputs settings Parameter menuT1 Input On/off Return airT3 Input On/off. Menu structure as T4 Input Menu 6 Digital inputs settings Digital inputs setting On/offMenu 7 Compartment settings Compartment On/offMenu Alarm settings Alarm group On/offMenu 9 Printer settings Delivery ticket settingsMenu 10 General settings Temperature units CMenu 11 Communication settings 10.2 Sample rate Minutes11.2 COM2 RS-232 port setting Accessories and spare parts Part Number Designation62-61138-20 02/03 Enclosure a Enclosure B Replace paper roll Enclosure C Factory settings Temperature inputs Factory settingCompartment settings Printer settingsEnclosure D Failure codes & Set correction factor procedure Enter PincodeEnclosure E Wiring diagrams Digital Inputs configuration example= GND = RXD = TXD = CTS