Chapter 3: Testing Performance
To test the time interval accuracy
To test the time interval accuracy
Testing the time interval accuracy does not check a specification, but does check the following:
•125 MHz oscillator
This test verifies that the 125 MHz timing acquisition synchronizing oscillator is operating within limits.
Equipment Required
Equipment | Critical Specifications | Recommended Model/Part |
Pulse Generator
Function Generator
SMA Cable
Adapter
BNC Test Connector, 6x2
200 MHz 2.5 ns pulse width, < 600 ps rise time Accuracy≤
8133A option 003
33250A
Set up the equipment
1Set up the logic analyzer:
a If you have not already done so, do the procedure “To set up the test equipment and the logic analyzer” on page 23.
b Exit and restart the Agilent Logic Analyzer applications to reinitialize the logic analyzer.
2Set up the pulse generator according to the following table.
Pulse Generator Setup
Timebase | Channel 1 | Trigger |
Mode: Ext | Mode: Square | Divide: Divide ⎟ 1 |
| Delay: 0.000 ns | Ampl: 0.50 V |
| High: | Ampl: 0.50 V |
| Low: | Offs: 0.00V |
| COMP: Disabled (LED off) |
|
|
|
|