Chapter 5: Troubleshooting
Troubleshooting the Agilent
eWhile the instrument is attempting to save the file to flexible disk, probe for digital signals on the flexible disk drive connector according to the following table.
Disk Drive Voltages
Pin | Signals | Pin No. | Signal |
1 | +5 V | 2 | INDEX |
3 | +5 V | 4 | DRIVE SELECT |
5 | +5 V | 6 | DISK CHANGE |
7 | NC | 8 | READY |
9 | NC | 10 | MOTOR ON |
11 | NC | 12 | DIRECTION SELECT |
13 | NC | 14 | STEP |
15 | 0 V | 16 | WRITE DATA |
17 | 0 V | 18 | WRITE GATE |
19 | 0 V | 20 | TRACK 00 |
21 | O V | 22 | WRITE PROTECT |
23 | 0 V | 24 | READ DATA |
25 | 0 V | 26 | SIDE ONE SELECT |
|
|
|
|
7Repeat steps 1 through 3 above. After the instrument turns off, unplug the instrument.
8Replace suspect disk drive if necessary, then reassemble the instrument.
To verify the CD-ROM
The