Alphabetical reference |
| Commands and Queries Reference | |
POLEN | enables the | XMP 2600 to accept signed value parameters. When | |
| enabled, the sign of VSET programmed values will control the output | ||
| polarity of the Power Module and voltage query replies will include a | ||
| negative sign when appropriate. |
| |
| Syntax | POLEN <int> |
|
| Parameters | a value of “0” disables and a value of “1” enables | |
|
| signed values. |
|
| Category | Settings | Type: Sequential |
POLEN? | return the current state of the Polarity Enable Flag. The result is a | ||
| decimal value in the range 0 to 1. |
| |
| Syntax | POLEN? |
|
| Parameters | none |
|
| Category | Settings | Type: Sequential |
POR | program the XMP 2600 behavior at Power On (Power On Retain). | ||
| Syntax | POR <int> |
|
| Parameters | a value of “0” will cause the XMP 2600 to initialize | |
|
| all Power Modules at Power On. A value of “1” will | |
|
| cause the XMP 2600 to | |
|
| Modules to the state they were at before power was | |
|
| turned off. |
|
| Category | System | Type: Sequential |
POR? | return the value of the Power On Retain Flag. | ||
| Syntax | POR? |
|
| Parameters | none |
|
| Category | System | Type: Sequential |
POWER? | return the XMP 2600 delivered | output power. The result is two | |
| decimal values: the first being the power delivered by the master | ||
| mainframe Power Modules and the second being the total output | ||
| power of all the Power Modules. |
| |
| Syntax | POWER? |
|
| Parameters | none |
|
| Category | Type: Sequential | |
PRIM? | return two decimal numbers, in the range 0 to 255. Each bit of the | ||
| reply represents a channel. A bit value of “1” means that the Power | ||
| Module is installed in the master mainframe (fed by the Primary). The | ||
| reply is a | ||
| Syntax | PRIM? |
|
| Parameters | none |
|
| Category | System | Type: Sequential |
56 | XMP 2600 Programming Manual rev. 1.1 |