Fisher FB1000-2 manual Safety Considerations

Page 3

Safety Considerations

Read and understand this manual completely before attempting to set up or use this instrument.

This equipment has been designed and tested to conform to IEC1010-1 safety standards, as applicable to laboratory instrumentation. This applies only to the FB1000 when used as specified in the documentation, in its intended applications, with Fisher Biotech approved electrophoresis apparatus only. Usage in any other manor may not provide similar performance or safety protection.

This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth ground is necessary for safe operation. Do not use any other AC line cord with this instrument.

The FB1000 is a high voltage power supply capable of generating dangerous levels of voltage and current during operation. Exercise caution when working around and with the electrical connections of this equipment. Always check electrical connectors, wires, and associated apparatus for any signs of wear or damage before using with this equipment. Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the FB1000 power supply.

The output of the FB1000 power supply is intended for connection to electrically isolated electrophoresis apparatus only. Use only with electrically isolated electrophoresis apparatus with minimum isolation of 1000V. Do not connect any terminal of the FB1000 output to earth ground. This may impair the safety protection provided by the equipment, or cause equipment damage.

The high voltage output of the FB1000 power supply takes some amount of time to decay when unloaded or lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching the power supply leads.

This equipment has a protective ground leakage current of 0.5mA at 120VAC, and 1mA at 230VAC. While significantly below the 3.5mA limitation for laboratory instrumentation, some applications and locations require a leakage current below 0.5mA. Check the specific requirements of your application before using this equipment.

This equipment is for indoor use only.

FB1000-2 MANUAL

Page 2

7608177 Rev. A 8/9/02

Image 3
Contents Operator’s Manual FB1000-2 Electrophoresis Power SupplyTable of Contents Fisher Biotech FB1000 Electrophoresis Power SupplySafety Considerations Introduction Unpacking the Power SupplyEnvironmental SpecificationsAC input Power DC output PowerGetting Started Using the Power SupplyConstant Power Operations Constant Voltage OperationsConstant Current Operations Programming Automatic Power-fail RestartTimed Operations Volt-Hour OperationsFB1000-2 Manual 100V / 200mA Prev CleaningMethod Next Steps Prev Edit Method 1 1/2 NextReplacing a fuse Troubleshooting and Error IndicationsFB1000 Warranty Statement Replacement PartsCompliance Table des matières Considérations de sécurité Caractéristiques Déballage de l’alimentation électriqueCourant de sortie CC SpécificationsCourant d’entrée CA EnvironnementUtilisation de l’alimentation électrique Mise en routeOpérations à intensité constante Opérations à tension constanteOpérations à puissance constante Opérations à durée déterminéeOpérations en volts-heures Programmation Redémarrage automatique en cas de coupure de courantFB1000-2 Manual 150W Add NettoyageMethod Next Steps PrevDépannage et indications d’erreur Remplacement d’un fusibleFisher Scientific Park Lane Pittsburgh, PA 15275 USA Garantie du FB1000Conformité Pièces de rechangeInhaltsverzeichnis Fisher Biotech FB1000 Elektrophorese-NetzteilSicherheitstechnische Hinweise Auspacken des Netzteils EinführungGleichstromausgangsleistung Technische DatenNetzeingangsleistung UmfeldbedingungenVerwenden des Netzteils EinstiegBetrieb bei Dauerstrom Betrieb bei DauerspannungBetrieb bei Dauerleistung Zeitgesteuerter BetriebVoltstundenbetrieb Programmierung Automatischer Neustart bei StromausfallMethod 1 1/2 Next 100V / 200mA Prev 150W Add 010 Edit 150W Hinzuf 010 Bearbeiten ReinigungAuswechseln einer Sicherung Fehlersuche und FehlermeldungFB1000 Garantieerklärung KompatibilitätErsatzteile Contenido Consideraciones de seguridad Desembalaje de la fuente de alimentación IntroducciónPotencia de salida de CC EspecificacionesPotencia de entrada de CA MedioambientalesUso de la fuente de alimentación Para empezarOperaciones a corriente constante Operaciones a voltaje constanteReinicio automático tras un fallo en la alimentación Operaciones a potencia constanteOperaciones temporizadas Operaciones en voltios-horaProgramación FB1000-2 Manual Método 1 1/2 Next LimpiezaMétodo Next Pasos PrevCondición y posible solución Diagnóstico e indicaciones de errorConformidad Conformidad EMC Compatibilidad electromagnética Declaración de garantía del FB1000Fisher Scientific Park Lane Pittsburgh, PA 15275, EE.UU Piezas de repuestoFB1000-2 Electrophoresis Power Supply Part No Rev a