OPERATING INSTRUCTIONS

STEP 4

Projecting

Rapidly rotate the control crank handle in the EXPOSE direction (counter-clockwise) to move the source assembly out of the exposure device to the radiographic focal position.

The control crank handle will stop turning when the source reaches the source stop. Do not use excessive force.

The odometer (if fitted) will indicate the approximate total distance traveled, 7 ft (2.1 m) for each guide tube section.

Set the brake to ON to prevent movement of the source assembly during the radiographic exposure.

Start timing the radiographic exposure from the moment the source assembly reaches the exposure head.

Survey meter readings observed during the projection operation should increase rapidly from background to a high level as the source emerges from the projector. Then readings should fall as the source moves out towards the focal point, fall sharply as the source enters a collimator (if used) and remain steady throughout the exposure.

Actual survey meter readings will depend on the source activity, distance, collimators and shielding. The sequence of changes should be observed and the readings noted.

During retraction of the source assembly from the exposure head to the exposure device, the sequence will reverse. The survey meter should indicate a continually increasing radiation level as the source assembly is retracted, then drop to background when the source assembly is stored in the exposure device.

During the radiographic exposure, use the survey meter to check the boundary dose rate, but spend as little time as possible in and near the restricted area to minimize personal exposure.

2.9

Image 20
Contents Operating and Maintenance Manual MAN-027-MayTime Page Front View Exposure Device Description Standard Source Assembly Model 880 Delta authorized contentsModel 880 Sigma authorized contents ApplicationsModel 880 Elite authorized contents Model 880 Omega authorized contentsSource output Selected attenuation dataSource Guide Tube Extension Control Specifications Complete Remote Control AssembliesSource Guide Tube with source stop Exposure Device Specifications Manufacturer Length Weight of Exposure Device Type BU CertificationWeight of Depleted Uranium Shield Primary Application WidthMaintenance Requirements Device Operating Temperature RangeSource Assembly ConstructionWarranty and Limitation of Liability Job site safety precautions Instruments Restricted AreaDistance ShieldingExposure Device Survey Source Guide Tube projection sheath LayoutSimultaneously pull and rotate Remote Control Conduit Layout Unlock the plunger lock with the key Important Warning Checks Before Exposure of the Source 1Unlocking Projecting Locking the Exposure Device RetractingConfirmatory Survey Dismantling equipment Remote control unit Source guide tubesSurveying StorageOperational Features Source Assembly IntroductionSource assembly and remote control cable connectors Exposure DeviceDaily Inspection Instructions Transit Mode Exposure ModeDaily Inspection of the System Daily Inspection of the Exposure DeviceDaily Inspection of the Remote Controls Daily Inspection of the Source Guide TubesRemote Control Inspection Important WarningControl Conduit Inspection Control Cable InspectionImportant Reminder Maintenance Instructions Routine Quarterly Maintenance Requirements Remote Controls Radiographic Exposure DeviceMisconnect Test Routine Quarterly Maintenance Records Annual Maintenance RequirementsLeak Testing Leak Test of the Radioactive Sealed Source Leak Test for Depleted Uranium DUAnnual Maintenance for the Model 880 Series Exposure Devices Model 880 Series Annual Maintenance Requirements Maintenance Instructions Maintenance Instructions Maintenance Instructions Annual Maintenance for the Remote Controls Remote Control Maintenance RequirementsMaintenance Instructions Maintenance Instructions Re-assembly of the Remote Control Unit Quarterly and Annual Maintenance for Source Guide Tubes Source Guide Tube MaintenanceMaintenance Instructions Annual Maintenance Records Considerations of Safety Class a ItemsRear Plate Assembly Rear Plate Assembly Front Plate Assembly Front Plate Assembly Control Assemblies Source Guide Tubes Control Assemblies 222 Transport Package Sample Transportation InstructionsReceipt of Radioactive Material Shipment of Radioactive Material Maximum Radiation Sample Transportation Instructions Sample Transportation Instructions Shipment of Empty Uranium Shielded Containers Sample Transportation Instructions Carriage of Radioactive Material Sample Transportation Instructions Radioactive material transport checklist Area alarm Alarm ratemeterApparatus for industrial gamma radiography Automatic securing mechanismRemote control Exposure headLocked position Maximum ratingWorking position Source holderSource changers Emergencies TrainingAccess to Restricted Areas United States RegulationsEU Regulations Personnel MonitoringDisposal Abc