MAINTENANCE INSTRUCTIONS

7Remove the front plate containing the outlet port mechanism from the empty Model 880 device by removing the four 5/16-18 x 1½ in socket button-head screws using the tamperproof tool bit mounted in a ratchet.

After removing the outlet port mechanism, clean the ‘S’ tube of the exposure device by pushing cloth swabs wetted with solvent through the ‘S’ tube until they come out clean. Use a dry cloth swab to remove any residual solvent from the ‘S’ tube after cleaning.

Clean the exterior of the Model 880 using a mild detergent solution to remove all dirt and grime. Visually inspect the exposure device’s stainless steel end plates for weld failures (cracks, etc.) on the locking mechanism and outlet port ends. Ensure the handle, the bottom contact surfaces and sides comprising the plastic jacket are intact. Check the bottom surfaces of the jacket to ensure the contact area is not excessively worn allowing contact of the stainless steel body with a flat work surface. Additionally, if dents to the device’s body or flanges due to accidental drops are found during this inspection, these repairs must be performed at a QSA Global service center.

Ensure that all of the information contained on the Model 880 label is legible. The trefoil and the warning, ‘Caution or Danger, Radioactive Material’, must be legible at a distance of 3 ft (approximately 1 m). The model number, serial number and Type B certification number (or Type A Specification identification) must also be legible. For label replacement, remove the old label from the exposure device by using a number 30 drill-bit mounted in a hand drill. Remove the rivet heads used to fasten the label to the stainless steel body by drilling through the rivet head just enough to allow the rivet head to rotate freely. Remove all rivet heads to remove label. Install the replacement label, andl and use a pop-rivet gun to secure the new 1/8 in x 3/16 in stainless steel rivets.

Disassemble the front plate (outlet port mechanism) by removing the set screw (or roll pin) from the outlet port cover. Unscrew the two socket head cap screws from the pivot disk. Remove and discard the two compression springs from the outlet port mechanism and replace with new springs.

Clean all parts in solvent using a brush to dislodge all dust and dirt. Dry all components thoroughly. Do not lubricate, leave all components completely dry.

Inspect for wear and burrs on the brass slider and rotor surfaces. Verify the tungsten port shield is not loose within the rotor. If the port shield is loose, remove the set screw, apply VibratiteTM thread sealant and retighten the set screw against the tungsten port shield. If the port shield uses a roll pin, remove the roll pin and replace. Inspect the front plate’s outlet port for wear and burrs where the source guide tube bayonet fitting engages.

Install new compression springs and assemble the outlet port mechanism. Apply VibratiteTM thread sealant to the socket head cap screws before tightening. Install a new set screw to the outlet port cover.

After assembly of the front plate, perform the following safety function tests using a bayonet source guide tube fitting. Pull the outlet port cover and rotate clockwise 90 degrees. Movement should be smooth and limited to a clockwise 90 degree turn. Insert a bayonet fitting into the outlet port and rotate 90 degrees in a counter-clockwise direction. Insertion and rotation of the bayonet fitting should be smooth and without snags. Rotate the outlet port cover from a 3 o’clock position to a 5 o’clock position. This operation moves the rotor from the port shield position to a pass through section of the rotor. Reverse the operation to disengage the bayonet fitting. Repeat the function test three times to ensure smooth operation.

4.9

Image 40
Contents Operating and Maintenance Manual MAN-027-MayTime Page Front View Exposure Device Description Standard Source Assembly Model 880 Delta authorized contentsModel 880 Sigma authorized contents ApplicationsModel 880 Elite authorized contents Model 880 Omega authorized contentsSource output Selected attenuation dataSource Guide Tube with source stop Control Specifications Complete Remote Control AssembliesSource Guide Tube Extension Exposure Device Specifications Manufacturer Length Weight of Exposure Device Type BU CertificationWeight of Depleted Uranium Shield Primary Application WidthMaintenance Requirements Device Operating Temperature RangeSource Assembly ConstructionWarranty and Limitation of Liability Job site safety precautions Instruments Restricted AreaDistance ShieldingExposure Device Survey Source Guide Tube projection sheath LayoutSimultaneously pull and rotate Remote Control Conduit Layout Unlock the plunger lock with the key Important Warning Checks Before Exposure of the Source 1Unlocking Projecting Confirmatory Survey RetractingLocking the Exposure Device Dismantling equipment Remote control unit Source guide tubesSurveying StorageOperational Features Source Assembly IntroductionSource assembly and remote control cable connectors Exposure DeviceDaily Inspection Instructions Transit Mode Exposure ModeDaily Inspection of the System Daily Inspection of the Exposure DeviceDaily Inspection of the Remote Controls Daily Inspection of the Source Guide TubesRemote Control Inspection Important WarningControl Conduit Inspection Control Cable InspectionImportant Reminder Maintenance Instructions Routine Quarterly Maintenance Requirements Remote Controls Radiographic Exposure DeviceMisconnect Test Routine Quarterly Maintenance Records Annual Maintenance RequirementsLeak Testing Leak Test of the Radioactive Sealed Source Leak Test for Depleted Uranium DUAnnual Maintenance for the Model 880 Series Exposure Devices Model 880 Series Annual Maintenance Requirements Maintenance Instructions Maintenance Instructions Maintenance Instructions Annual Maintenance for the Remote Controls Remote Control Maintenance RequirementsMaintenance Instructions Maintenance Instructions Re-assembly of the Remote Control Unit Quarterly and Annual Maintenance for Source Guide Tubes Source Guide Tube MaintenanceMaintenance Instructions Annual Maintenance Records Considerations of Safety Class a ItemsRear Plate Assembly Rear Plate Assembly Front Plate Assembly Front Plate Assembly Control Assemblies Source Guide Tubes Control Assemblies 222 Transport Package Sample Transportation InstructionsReceipt of Radioactive Material Shipment of Radioactive Material Maximum Radiation Sample Transportation Instructions Sample Transportation Instructions Shipment of Empty Uranium Shielded Containers Sample Transportation Instructions Carriage of Radioactive Material Sample Transportation Instructions Radioactive material transport checklist Area alarm Alarm ratemeterApparatus for industrial gamma radiography Automatic securing mechanismRemote control Exposure headLocked position Maximum ratingSource changers Source holderWorking position Emergencies TrainingAccess to Restricted Areas United States RegulationsEU Regulations Personnel MonitoringDisposal Abc