Emerson Process Management MSL600 manual Noise Alrm Lev P622, Parameter values metric

Page 38

 

Installation, Operation & Maintenance Manual

MSL600

IP262/Z0, Rev. AB

February 2012

 

 

7.9.2.2Noise Alrm Lev – P622

The unit automatically monitors the background electronic noise floor as a % of display width, see parameter D834. If this level exceeds the Noise Alarm Lev value, programmed as a % of display width then an alarm can be indicated by allocating an output in P552, Noise Alarm.

The default value is 25%.

Parameter values:

Default

Min

Max

25

0

100

7.9.2.3Top Blanking – P623

This is the minimum operating range for the transducer in metres (feet – imperial). This parameter allows the user to program a region in front of the transducer which is not used by the system.

This is useful if there are any fixed obstructions in front of the transducer which cannot be moved or otherwise avoided.

This parameter is also useful when programming the industrial algorithm for avoiding floating or settling debris in front of the transducer.

The default value is 0.3 metres / 1 foot.

Parameter values (metric)

 

Parameter values (imperial)

Default

Min

Max

Default

Min

Max

0.300

0.100

7.000

1.00

0.33

23.00

7.9.2.4Damping – P624

It is unlikely that the blanket level will change by a large amount on a short time scale. The settling process is typically fairly slow. If the system should want to respond to noise or stray echoes from bubbles or debris this parameter, which is programmed as a time measured in seconds damps out rapid fluctuations in the measured sludge blanket level.

The default value is 60 seconds.

Parameter values:

Default

Min

Max

60

0

9999

7.9.2.5Upr S-nat Temp – P625

This is the value used for temperature compensation if the measured supernatant temperature should exceed this value. Valid entries are between –35 and +85°C (-30 to 185 deg F).

The default is +40°C (105 deg F)

Parameter values (metric)

 

Parameter values (imperial)

Default

Min

Max

Default

Min

Max

40

-35

+85

105

-30

185

Page 38

Image 38
Contents Installation, Operation & Maintenance Manual MSL600Contents MSL600 12.1 12.2Before Operating this Device the Manual should be Read Fully About the MSL600 About this ManualHuman Machine Interface Important MSL600 featuresSelf-cleaning Mounting bracketAs a system LocationRotating bridge Application planningCircular clarifiers Rectangular clarifiers Static bridgeLocation InstallationPreliminary checks Mounting bracket attachmentInstalling the bracket on the bridge Attaching the MSL600MSL600 mounting Power and other electrical connections Transducer connectionsExternal connection terminals Cable extension Programming DisplayContrast viewing angle control Keypad and Menu NavigationSecurity On-line/Off-lineQuick start parameters Tank depth Bottom referenceTransducer self-cleaning Determination of tank depthCurrent output adjust MA current output parametersTransducer out-of-water time AlarmRelay parameters Digital input Main menu item Detailed operationMenu structure Menu Item DescriptionScrolling Direct accessParameter editing Main menu parameters Installation parametersApplication parameters Start On P250 De-sludge parametersDescription P240 Stop On P251Interval P254 Stop If P252Start Time P253 Max Retries P257Cleaning parameters MA output parametersAlarm Action P402 4 0/4-20mA P403Table B Relay Parameter numbers Relay RL1 mode P410 and RL2 mode P420Alarm mode relay RL1 and RL2 parameters P540 to De-sludge mode relay RL1 and RL2 Fault mode relay RL1 and RL2 parameters P560 to P570Set point mode relay RL1 and RL2 On mode relay RL1 and RL2Relay overrides Min ON, Max ON, and Min OFF Display parameters HMIRelay On and Off Points HMI upper, middle, and lower display optionsClear Profile Backlight On/Off P575Clear Trend Log Interval P590 Low Mem Alarm P593Logging parameters Fast log P5911 Alg. Select P620 Parameter values Default Min Max 128 1024Engineering parameters Common algorithm parameters P621 to P629Default Min Max +85 105 185 Parameter values Default Min Max 100 Top Blanking P623Parameter values Default Min Max 9999 Upr S-nat Temp P625 Noise Alrm Lev P622Lwr S-nat Temp P626 Default Min Max +85 185 SoS constant P627Default metric m/s Bottom Blanking P628Significance P630 Parameter values Default Min Max Lost Echo Rnge P633Municipal 1 parameters P630 to P637 Sharp Echo Reject P631Parameter values Default Min Max 255 Parameter values Default Min Max Dirty Mult P636Parameter values Default Min Max Lost Echo Time P637 Noise SD Mult P635Threshld Offset P640 Parameter values Default Min Max 100Municipal 2 parameter P640 Threshold level = Threshld Offset P640 + Noise Level D834Auto Cycle function Load Defaults functionSystem parameters Display testSettings parameters P730 to P737 and Base Units selection Comms parameters P710 to P716PIN parameter P740 Fixed parameters D750 to D753Readings and diagnostics parameters Readings parameters D800 to D834Diagnostics parameters D835 to D852 Min Xdr Temp D852 Speed of Sound D850Max Xdr Temp D851 Technical reference MaintenanceTroubleshooting Transmit pulse and echo processingProgramming chart Sub-menu Level Parameter Title Par No Main Menu OptionProgramming chart … Direct Model No Specification12.1 MSL600 Outputs12.2 MSL603 transducer WeightAdditional specification notes Dimensional drawings MSL600 Bridge mounted unitMSL600 and MSL603 assembly Fitting instructions MSL603 transducer cable extensionParts required MSL600 MSL600 Emerson Process Management Rosemount Measurement