7.9.2.6Lwr S-nat Temp – P626

This value is used for temperature compensation if the measured supernatant temperature is below this value.

The lower temperature (P626) must always be less than the upper (P625). Valid entries are between –40 and +85°C (-40 to 185 deg F).

The default is -5°C (25 deg F)

Parameter values (metric)

 

Parameter values (imperial)

Default

Min

Max

Default

Min

Max

-5

-40

+85

25

-40

185

7.9.2.7SoS constant - P627

The speed of sound at 0oC (32 deg F) for a particular supernatant can be programmed in m/s (ft/s imperial).

Default (metric m/s)

Default (imperial ft/s)

1402.0

4599.7

7.9.2.8Bottom Blanking – P628

This parameter allows the user to program a region close to the bottom of the tank which is not used by the system for normal blanket level measurement.

This is useful if there are any fixed obstructions in front of the transducer close to the bottom of the tank which cannot be moved or otherwise avoided.

The default value is 0.0 metres / 0 foot

Parameter values (metric)

 

Parameter values (imperial)

Default

Min

Max

Default

Min

Max

0.000

0.000

9.999

0.00

0.00

33.00

7.9.2.9Level Offset – P629

This parameter is used to provide an offset to the normal displayed sludge blanket level (D800). The value may be entered as a positive or negative value to provide the required offset to the normal displayed sludge blanket level.

Default is 0.0.

 

 

 

 

 

Parameter values (metric)

 

Parameter values (imperial)

Default

Min

Max

Default

Min

Max

0.0

-0.99

0.99

0.0

-3.0

3.0

Page 39

Image 39
Contents MSL600 Installation, Operation & Maintenance ManualContents MSL600 12.2 12.1Before Operating this Device the Manual should be Read Fully About this Manual About the MSL600Mounting bracket Important MSL600 featuresSelf-cleaning Human Machine InterfaceLocation As a systemApplication planning Circular clarifiersRotating bridge Static bridge Rectangular clarifiersMounting bracket attachment InstallationPreliminary checks LocationAttaching the MSL600 Installing the bracket on the bridgeMSL600 mounting Transducer connections Power and other electrical connectionsExternal connection terminals Cable extension Display ProgrammingKeypad and Menu Navigation Contrast viewing angle controlOn-line/Off-line SecurityTank depth Bottom reference Quick start parametersDetermination of tank depth Transducer self-cleaningMA current output parameters Current output adjustAlarm Relay parametersTransducer out-of-water time Digital input Menu Item Description Detailed operationMenu structure Main menu itemDirect access Parameter editingScrolling Installation parameters Application parametersMain menu parameters Stop On P251 De-sludge parametersDescription P240 Start On P250Max Retries P257 Stop If P252Start Time P253 Interval P254MA output parameters Cleaning parameters4 0/4-20mA P403 Alarm Action P402Relay RL1 mode P410 and RL2 mode P420 Alarm mode relay RL1 and RL2 parameters P540 toTable B Relay Parameter numbers On mode relay RL1 and RL2 Fault mode relay RL1 and RL2 parameters P560 to P570Set point mode relay RL1 and RL2 De-sludge mode relay RL1 and RL2HMI upper, middle, and lower display options Display parameters HMIRelay On and Off Points Relay overrides Min ON, Max ON, and Min OFFBacklight On/Off P575 Clear TrendClear Profile Fast log P591 Low Mem Alarm P593 Logging parameters Log Interval P590Common algorithm parameters P621 to P629 Parameter values Default Min Max 128 1024Engineering parameters 1 Alg. Select P620Noise Alrm Lev P622 Parameter values Default Min Max 100 Top Blanking P623Parameter values Default Min Max 9999 Upr S-nat Temp P625 Default Min Max +85 105 185Bottom Blanking P628 Default Min Max +85 185 SoS constant P627Default metric m/s Lwr S-nat Temp P626Sharp Echo Reject P631 Parameter values Default Min Max Lost Echo Rnge P633Municipal 1 parameters P630 to P637 Significance P630Noise SD Mult P635 Parameter values Default Min Max Dirty Mult P636Parameter values Default Min Max Lost Echo Time P637 Parameter values Default Min Max 255Threshold level = Threshld Offset P640 + Noise Level D834 Parameter values Default Min Max 100Municipal 2 parameter P640 Threshld Offset P640Display test Load Defaults functionSystem parameters Auto Cycle functionComms parameters P710 to P716 Settings parameters P730 to P737 and Base Units selectionFixed parameters D750 to D753 PIN parameter P740Readings parameters D800 to D834 Readings and diagnostics parametersDiagnostics parameters D835 to D852 Speed of Sound D850 Max Xdr Temp D851Min Xdr Temp D852 Transmit pulse and echo processing MaintenanceTroubleshooting Technical referenceSub-menu Level Parameter Title Par No Main Menu Option Programming chartProgramming chart … Direct Outputs Specification12.1 MSL600 Model NoWeight Additional specification notes12.2 MSL603 transducer MSL600 Bridge mounted unit Dimensional drawingsMSL600 and MSL603 assembly MSL603 transducer cable extension Parts requiredFitting instructions MSL600 MSL600 Emerson Process Management Rosemount Measurement