Emerson Process Management MSL600 manual Cleaning parameters, MA output parameters

Page 30

 

Installation, Operation & Maintenance Manual

MSL600

IP262/Z0, Rev. AB

February 2012

 

 

7.4 Cleaning parameters

The following parameter relates to the cleaning function.

Note: A full menu structure is shown in section 11.

7.4.1Start On - P260

Use this parameter to select how the cleaning cycle starts. The selectable options are:

￿AutoClean

￿Manual 5s

￿Off

The default is “AutoClean”.

7.4.2Interval - P264

Use this parameter to programme the interval between cleaning cycles (see also P444 - cleaning duration). This parameter is programmed in hours and minutes.

Default

Min

Max

1:00

0:10

1:39

7.4.3Cleaning Time - P444

The duration of the cleaning cycle can be programmed in seconds

Default

Min

Max

0:10

0:10

0:20

7.5 mA output parameters

The following parameters are used to program the mA current output.

7.5.1Lower Range Val - P400

This is the sludge blanket level in metres (or feet if Imperial version) for the minimum mA output, either 0 or 4mA as programmed using P403.

Parameter values (Metric)

 

Parameter values

(Imperial)

Default

Min

Max

Default

Min

Max

0.000m

0.000m

7.000m

0.00 ft

0.00 ft

23.00 ft

7.5.2Upper Range Val - P401

This is the sludge blanket level in metres (or feet if Imperial version) for the maximum mA output. This value is by default set to "Auto".

Parameter values (Metric)

 

Parameter values

(Imperial)

Default

Min

Max

Default

Min

Max

Auto

0.000m

7.000m

Auto

0.00 ft

23.00 ft

Note: In the case of the default setting the upper limit is defined as being at the programmed transducer blanking distance P623 from the front of the transducer face.

Page 30

Image 30
Contents Installation, Operation & Maintenance Manual MSL600Contents MSL600 12.1 12.2Before Operating this Device the Manual should be Read Fully About the MSL600 About this ManualHuman Machine Interface Important MSL600 featuresSelf-cleaning Mounting bracketAs a system LocationApplication planning Circular clarifiersRotating bridge Rectangular clarifiers Static bridgeLocation InstallationPreliminary checks Mounting bracket attachmentInstalling the bracket on the bridge Attaching the MSL600MSL600 mounting Power and other electrical connections Transducer connectionsExternal connection terminals Cable extension Programming DisplayContrast viewing angle control Keypad and Menu NavigationSecurity On-line/Off-lineQuick start parameters Tank depth Bottom referenceTransducer self-cleaning Determination of tank depthCurrent output adjust MA current output parametersAlarm Relay parametersTransducer out-of-water time Digital input Main menu item Detailed operationMenu structure Menu Item DescriptionDirect access Parameter editingScrolling Installation parameters Application parametersMain menu parameters Start On P250 De-sludge parametersDescription P240 Stop On P251Interval P254 Stop If P252Start Time P253 Max Retries P257Cleaning parameters MA output parametersAlarm Action P402 4 0/4-20mA P403Relay RL1 mode P410 and RL2 mode P420 Alarm mode relay RL1 and RL2 parameters P540 toTable B Relay Parameter numbers De-sludge mode relay RL1 and RL2 Fault mode relay RL1 and RL2 parameters P560 to P570Set point mode relay RL1 and RL2 On mode relay RL1 and RL2Relay overrides Min ON, Max ON, and Min OFF Display parameters HMIRelay On and Off Points HMI upper, middle, and lower display optionsBacklight On/Off P575 Clear TrendClear Profile Log Interval P590 Low Mem Alarm P593Logging parameters Fast log P5911 Alg. Select P620 Parameter values Default Min Max 128 1024Engineering parameters Common algorithm parameters P621 to P629Default Min Max +85 105 185 Parameter values Default Min Max 100 Top Blanking P623Parameter values Default Min Max 9999 Upr S-nat Temp P625 Noise Alrm Lev P622Lwr S-nat Temp P626 Default Min Max +85 185 SoS constant P627Default metric m/s Bottom Blanking P628Significance P630 Parameter values Default Min Max Lost Echo Rnge P633Municipal 1 parameters P630 to P637 Sharp Echo Reject P631Parameter values Default Min Max 255 Parameter values Default Min Max Dirty Mult P636Parameter values Default Min Max Lost Echo Time P637 Noise SD Mult P635Threshld Offset P640 Parameter values Default Min Max 100Municipal 2 parameter P640 Threshold level = Threshld Offset P640 + Noise Level D834Auto Cycle function Load Defaults functionSystem parameters Display testSettings parameters P730 to P737 and Base Units selection Comms parameters P710 to P716PIN parameter P740 Fixed parameters D750 to D753Readings and diagnostics parameters Readings parameters D800 to D834Diagnostics parameters D835 to D852 Speed of Sound D850 Max Xdr Temp D851Min Xdr Temp D852 Technical reference MaintenanceTroubleshooting Transmit pulse and echo processingProgramming chart Sub-menu Level Parameter Title Par No Main Menu OptionProgramming chart … Direct Model No Specification12.1 MSL600 OutputsWeight Additional specification notes12.2 MSL603 transducer Dimensional drawings MSL600 Bridge mounted unitMSL600 and MSL603 assembly MSL603 transducer cable extension Parts requiredFitting instructions MSL600 MSL600 Emerson Process Management Rosemount Measurement