Emerson Process Management MSL600 manual Quick start parameters, Tank depth Bottom reference

Page 20

 

Installation, Operation & Maintenance Manual

MSL600

IP262/Z0, Rev. AB

February 2012

 

 

5. Quick start parameters

The system leaves the factory with default values in all the parameters. These values, after installation, allow immediate operation without any further programming. However, it is recommended for best operation that the following parameters should be programmed.

In order to get meaningful measurements from the MSL600 in the minimum time only the following parameters need to be adjusted.

5.1 Tank depth - Bottom reference

An accurate knowledge of the tank depth is a key parameter for the system.

The tank depth is the distance from the front face of the transducer to the bottom of the tank directly under the transducer [see Figure (10)]. Where the tank bottom slopes, as is often the case, the sludge blanket level, Xm indicated by the system is therefore relative to this zero reference level.

Tank depth

Sludge blanket interface Xm

0m tank depth

Figure (10): Tank depth parameter

The maximum tank depth that the system will measure is 7.000m (23.00 feet), the minimum is 1.000m (3.3 feet), do not confuse this with the transducer blanking distance which is discussed elsewhere and is 0.3m (1 foot) by default.

The tank depth value is stored in the system and used in subsequent calculations. The factory default is 7.000m (23.00 feet), corresponding to the maximum depth of tank for which the system is specified.

When the unit is powered on for the first time the system makes measurements using this default. It will be obvious that until the user enters a measured value certain functions, such as current output and displayed blanket level will not be correct for the installation.

Page 20

Image 20
Contents Installation, Operation & Maintenance Manual MSL600Contents MSL600 12.1 12.2Before Operating this Device the Manual should be Read Fully About the MSL600 About this ManualImportant MSL600 features Self-cleaningHuman Machine Interface Mounting bracketAs a system LocationRotating bridge Application planningCircular clarifiers Rectangular clarifiers Static bridgeInstallation Preliminary checksLocation Mounting bracket attachmentInstalling the bracket on the bridge Attaching the MSL600MSL600 mounting Power and other electrical connections Transducer connectionsExternal connection terminals Cable extension Programming DisplayContrast viewing angle control Keypad and Menu NavigationSecurity On-line/Off-lineQuick start parameters Tank depth Bottom referenceTransducer self-cleaning Determination of tank depthCurrent output adjust MA current output parametersTransducer out-of-water time AlarmRelay parameters Digital input Detailed operation Menu structureMain menu item Menu Item DescriptionScrolling Direct accessParameter editing Main menu parameters Installation parametersApplication parameters De-sludge parameters Description P240Start On P250 Stop On P251Stop If P252 Start Time P253Interval P254 Max Retries P257Cleaning parameters MA output parametersAlarm Action P402 4 0/4-20mA P403Table B Relay Parameter numbers Relay RL1 mode P410 and RL2 mode P420Alarm mode relay RL1 and RL2 parameters P540 to Fault mode relay RL1 and RL2 parameters P560 to P570 Set point mode relay RL1 and RL2De-sludge mode relay RL1 and RL2 On mode relay RL1 and RL2Display parameters HMI Relay On and Off PointsRelay overrides Min ON, Max ON, and Min OFF HMI upper, middle, and lower display optionsClear Profile Backlight On/Off P575Clear Trend Low Mem Alarm P593 Logging parametersLog Interval P590 Fast log P591Parameter values Default Min Max 128 1024 Engineering parameters1 Alg. Select P620 Common algorithm parameters P621 to P629Parameter values Default Min Max 100 Top Blanking P623 Parameter values Default Min Max 9999 Upr S-nat Temp P625Default Min Max +85 105 185 Noise Alrm Lev P622Default Min Max +85 185 SoS constant P627 Default metric m/sLwr S-nat Temp P626 Bottom Blanking P628Parameter values Default Min Max Lost Echo Rnge P633 Municipal 1 parameters P630 to P637Significance P630 Sharp Echo Reject P631Parameter values Default Min Max Dirty Mult P636 Parameter values Default Min Max Lost Echo Time P637Parameter values Default Min Max 255 Noise SD Mult P635Parameter values Default Min Max 100 Municipal 2 parameter P640Threshld Offset P640 Threshold level = Threshld Offset P640 + Noise Level D834Load Defaults function System parametersAuto Cycle function Display testSettings parameters P730 to P737 and Base Units selection Comms parameters P710 to P716PIN parameter P740 Fixed parameters D750 to D753Readings and diagnostics parameters Readings parameters D800 to D834Diagnostics parameters D835 to D852 Min Xdr Temp D852 Speed of Sound D850Max Xdr Temp D851 Maintenance TroubleshootingTechnical reference Transmit pulse and echo processingProgramming chart Sub-menu Level Parameter Title Par No Main Menu OptionProgramming chart … Direct Specification 12.1 MSL600Model No Outputs12.2 MSL603 transducer WeightAdditional specification notes Dimensional drawings MSL600 Bridge mounted unitMSL600 and MSL603 assembly Fitting instructions MSL603 transducer cable extensionParts required MSL600 MSL600 Emerson Process Management Rosemount Measurement