Emerson Process Management MSL600 manual Municipal 2 parameter P640, Threshld Offset P640

Page 42

 

Installation, Operation & Maintenance Manual

MSL600

IP262/Z0, Rev. AB

February 2012

 

 

7.9.4Municipal 2 parameter P640

Only one specific parameter, P640, needs to be adjusted when the Municipal 2 algorithm is selected.

Industrial blankets are typically much denser and more substantial than municipal sludge found in a conventional wastewater plant. For this reason, the echo profile from an industrial sludge typically only displays one distinct echo.

7.9.4.1Threshld Offset – P640

The echo from the blanket will produce a signal which is visible above the electronic noise floor on the echo profile display.

The threshold offset parameter is the offset of the threshold which the user wishes to set above the noise floor. Both the noise level, D834 and the threshold offset are expressed as a % of full screen.

The noise level can be viewed as a diagnostic parameter from the monitor menu. The first point in the echo profile which exceeds the threshold level is determined by the algorithm to be at the blanket level.

Remember that;

Threshold level = Threshld Offset (P640) + Noise Level (D834)

When setting this parameter, remember that the echo amplitude can get larger or smaller over a period of time during normal operation of the system. This is because the consistency of the sludge varies with time and position in the settling tank as the bridge rotates over it.

Floating debris close to the transducer can give echoes which will exceed the threshold level. It is suggested that in this case the blanking distance, P623 be extended to include this region in the echo profile.

The default value is 30%.

Parameter values:

Default

Min

Max

30

0

100

Page 42

Image 42
Contents Installation, Operation & Maintenance Manual MSL600Contents MSL600 12.1 12.2Before Operating this Device the Manual should be Read Fully About the MSL600 About this ManualHuman Machine Interface Important MSL600 featuresSelf-cleaning Mounting bracketAs a system LocationApplication planning Circular clarifiersRotating bridge Rectangular clarifiers Static bridgeLocation InstallationPreliminary checks Mounting bracket attachmentInstalling the bracket on the bridge Attaching the MSL600MSL600 mounting Power and other electrical connections Transducer connectionsExternal connection terminals Cable extension Programming DisplayContrast viewing angle control Keypad and Menu NavigationSecurity On-line/Off-lineQuick start parameters Tank depth Bottom referenceTransducer self-cleaning Determination of tank depthCurrent output adjust MA current output parametersAlarm Relay parametersTransducer out-of-water time Digital input Main menu item Detailed operationMenu structure Menu Item DescriptionDirect access Parameter editingScrolling Installation parameters Application parametersMain menu parameters Start On P250 De-sludge parametersDescription P240 Stop On P251Interval P254 Stop If P252Start Time P253 Max Retries P257Cleaning parameters MA output parametersAlarm Action P402 4 0/4-20mA P403Relay RL1 mode P410 and RL2 mode P420 Alarm mode relay RL1 and RL2 parameters P540 toTable B Relay Parameter numbers De-sludge mode relay RL1 and RL2 Fault mode relay RL1 and RL2 parameters P560 to P570Set point mode relay RL1 and RL2 On mode relay RL1 and RL2Relay overrides Min ON, Max ON, and Min OFF Display parameters HMIRelay On and Off Points HMI upper, middle, and lower display optionsBacklight On/Off P575 Clear TrendClear Profile Log Interval P590 Low Mem Alarm P593Logging parameters Fast log P5911 Alg. Select P620 Parameter values Default Min Max 128 1024Engineering parameters Common algorithm parameters P621 to P629Default Min Max +85 105 185 Parameter values Default Min Max 100 Top Blanking P623Parameter values Default Min Max 9999 Upr S-nat Temp P625 Noise Alrm Lev P622Lwr S-nat Temp P626 Default Min Max +85 185 SoS constant P627Default metric m/s Bottom Blanking P628Significance P630 Parameter values Default Min Max Lost Echo Rnge P633Municipal 1 parameters P630 to P637 Sharp Echo Reject P631Parameter values Default Min Max 255 Parameter values Default Min Max Dirty Mult P636Parameter values Default Min Max Lost Echo Time P637 Noise SD Mult P635Threshld Offset P640 Parameter values Default Min Max 100Municipal 2 parameter P640 Threshold level = Threshld Offset P640 + Noise Level D834Auto Cycle function Load Defaults functionSystem parameters Display testSettings parameters P730 to P737 and Base Units selection Comms parameters P710 to P716PIN parameter P740 Fixed parameters D750 to D753Readings and diagnostics parameters Readings parameters D800 to D834Diagnostics parameters D835 to D852 Speed of Sound D850 Max Xdr Temp D851Min Xdr Temp D852 Technical reference MaintenanceTroubleshooting Transmit pulse and echo processingProgramming chart Sub-menu Level Parameter Title Par No Main Menu OptionProgramming chart … Direct Model No Specification12.1 MSL600 OutputsWeight Additional specification notes12.2 MSL603 transducer Dimensional drawings MSL600 Bridge mounted unitMSL600 and MSL603 assembly MSL603 transducer cable extension Parts requiredFitting instructions MSL600 MSL600 Emerson Process Management Rosemount Measurement