Mode 01 (~LV01)
Mode 01 I/O operation
1.System will operate in Edge Sync or Envelope sync mode (~LX#)
2.Sync polarity programmable (~LP#)
3.Use ~PR0210 for proper port initialization.
4.Port 1 will go active OFF (no current) on an error condition.
5.Port 2 will go active ON (sink current) on an error condition.
6.Pushing the reset button or power
7.Error conditions available are:
a.Partial Read (Programmable, ~LQ#, ~Lp#)
b.% decode (Programmable, ~LD##)
c.Bad Quiet Zone (~PB816xxx100)
d.ISO Defects grade (Programmable, ~PB806xxx000)
e.ISO Decodability (Programmable, ~PB802xxx100)
f.Symbol Contrast grade (Programmable, ~PB804xxx100)
g.No Read (if sync is received)
h.Number of codes per sync programmable (~LN##, ~LZ##)
i.Overall ISO Grade (Programmable, ~LAxx)
j.Symbology modulo check digit
Mode 01 LED Operation
1.LED1 will turn on if Outputs 1 and 2 go active due to ISO, contrast, or quiet zone failure
2.LED2 will turn on if Outputs 1 and 2 go active due to a partial or no read condition
Quick Check® SV Series User’s Guide | 6 - 31 |